Effect of a novel interspinous implant on lumbar spinal range of motion.

机译:新型棘突间植入物对腰椎运动范围的影响。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Interspinous devices have been introduced to provide a minimally invasive surgical alternative for patients with lumbar spinal stenosis or foraminal stenosis. Little is known however, of the effect of interspinous devices on intersegmental range of motion (ROM). The aim of this in vivo study was to investigate the effect of a novel minimally invasive interspinous implant, InSwing, on sagittal plane ROM of the lumbar spine using an ovine model. Ten adolescent Merino lambs underwent a destabilization procedure at the L1-L2 level simulating a stenotic degenerative spondylolisthesis (as described in our earlier work; Spine 15:571-576, 1990). All animals were placed in a side-lying posture and lateral radiographs were taken in full flexion and extension of the trunk in a standardized manner. Radiographs were repeated following the insertion of an 8-mm InSwing interspinous device at L1-L2, and again with the implant secured by means of a tension band tightened to 1 N/m around the L1 and L2 spinous processes. ROM was assessed in each of the three conditions and compared using Cobb's method. A paired t-test compared ROM for each of the experimental conditions (P < 0.05). After instrumentation with the InSwing interspinous implant, the mean total sagittal ROM (from full extension to full flexion) was reduced by 16% from 6.3 degrees to 5.3 +/- 2.7 degrees. The addition of the tension band resulted in a 43% reduction in total sagittal ROM to 3.6 +/- 1.9 degrees which approached significance. When looking at flexion only, the addition of the interspinous implant without the tension band did not significantly reduce lumbar flexion, however, a statistically significant 15% reduction in lumbar flexion was observed with the addition of the tension band (P = 0.01). To our knowledge, this is the first in vivo study radiographically showing the advantage of using an interspinous device to stabilize the spine in flexion. These results are important findings particularly for patients with clinical symptoms related to instable degenerative spondylolisthesis.
机译:棘突间装置已被引入以为患有腰椎管狭窄或椎间孔狭窄的患者提供微创手术替代方案。然而,人们很少了解棘突间装置对节间运动范围(ROM)的影响。这项体内研究的目的是使用绵羊模型研究新型微创棘突间植入物InSwing对腰椎矢状面ROM的影响。十只青春期的美利奴羊羔在L1-L2水平上进行了去稳定化过程,模拟了狭窄的退化性腰椎滑脱(如我们先前的工作所述; Spine 15:571-576,1990)。将所有动物以侧卧姿势放置,并以标准方式在躯干完全弯曲和伸展下拍摄侧位X线照片。在L1-L2处插入8毫米InSwing棘突间装置后,重复进行X线照片,并再次通过围绕L1和L2棘突拉紧至1 N / m的张力带固定植入物。在这三个条件下分别评估ROM,并使用Cobb方法进行比较。配对的t检验比较了每种实验条件下的ROM(P <0.05)。在使用InSwing棘突间植入物进行器械植入后,平均总矢状ROM(从完全伸展到完全屈曲)从6.3度降低了16%至5.3 +/- 2.7度。张力带的加入使总矢状ROM降低了43%,降至3.6 +/- 1.9度,已接近显着水平。仅看屈曲时,增加不带张力带的棘突间植入物不会显着降低腰部屈曲,但是,观察到在增加张力带的情况下,腰椎屈曲率统计学上显着降低了15%(P = 0.01)。据我们所知,这是第一个通过射线照相显示的体内研究,显示了使用棘突间装置稳定脊柱屈曲的优势。这些结果是重要的发现,特别是对于具有与不稳定的退行性腰椎滑脱相关的临床症状的患者。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号