Fluid-dynamic results of in vitro comparison of four pericardial bioprostheses implanted in small porcine aortic roots

机译:体外比较四个猪小动脉主根植入心包生物假体的流体动力学结果

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: Small-sized aortic bioprostheses may cause high postoperative gradients. In clinical practice, it is difficult to compare bioprostheses from different manufactures, owing to the discrepancy between the true size and the nominal size of the prosthesis and the inter-patient variability in aortic root characteristics. In vitro studies provide accurate data, and using a system in which it is possible to implant bioprostheses in a true aortic root should enable a fair comparison to be made. The present study compared the four most widely used pericardial stented bioprostheses from different manufacturers surgically implanted in small annulus, to detect any differences in their fluid-dynamic performance.
机译:目的:小型主动脉生物假体可能导致较高的术后梯度。在临床实践中,由于假体的真实尺寸和标称尺寸以及主动脉根部特征的患者间差异,因此很难比较不同制造商的生物假体。体外研究提供了准确的数据,并且使用可以在真实的主动脉根部植入生物假体的系统应该能够进行公平的比较。本研究比较了不同厂家生产的四种最广泛使用的心包支架式生物假体,这些假体通过小环植入手术,以检测其流体动力学性能的任何差异。