摘要

Expandable vertebral body replacement systems have been increasingly used for anterior stabilization of spine. We report a secondary collapse of an expandable vertebral body replacement system. This specific complication has not been reported in the literature so far. The most obvious reason for failure was insufficient tightening of a locking screw. This paper emphasizes the importance of correct technical application.
机译:可扩展的椎体置换系统已越来越多地用于脊柱的前稳定。我们报告了可膨胀椎体置换系统的继发性塌陷。迄今为止,尚未在文献中报道这种特定的并发症。失败的最明显原因是锁定螺钉的拧紧不足。本文强调正确技术应用的重要性。