Magnetic resonance-guided focused ultrasound surgery using an enhanced sonication technique in a pig muscle model.

机译:在猪肌肉模型中使用增强超声技术的磁共振引导聚焦超声手术。

获取原文
获取原文并翻译 | 示例
           

摘要

THE PURPOSE OF THIS STUDY: To evaluate the safety and efficacy of an enhanced magnetic resonance-guided focused ultrasound (MRgFUS) emission protocol that results in more extensive treatment by increasing the volume of each focal ablation using the same energy. MATERIALS AND METHODS: Six pigs were treated with an MRgFUS system combined with real-time MR, for imaging and temperature mapping, with 102 "enhanced" and 97 "regular" focal ablations performed on both buttock muscles. Real-time imaging, temperature mapping, and acoustic reflected spectrum data enabled immediate evaluation of the results. MR contrast-enhanced images and pathology examinations were used for confirmation. RESULTS: The location of the ablated volume by "enhanced" sonication is predictable, with a maximum possible shift of 6 mm toward, and 3 mm away, from the transducer. The ablated volume after enhanced sonication was, on average, 1.8 times larger than after a regular sonication of the same energy. Pathology results showed the same thermally induced damage patterns in the enhanced sonications and the regular sonications. CONCLUSION: Accelerated MRgFUS with enhanced sonication is a safe, controllable, and more effective tissue ablative modality than standard sonication. This new technology may significantly reduce the length of tumor ablation procedures. (Isn't the new technology you're talking about MRgFUS? If so, you don't need to repeat it at the end of this sentence.).
机译:这项研究的目的:评估增强的磁共振引导聚焦超声(MRgFUS)发射协议的安全性和有效性,该协议通过使用相同能量增加每次局灶消融的体积来导致更广泛的治疗。材料与方法:6头猪用MRgFUS系统结合实时MR进行处理,以进行成像和温度测绘,对两只臀部肌肉进行102次“增强”和97次“常规”局部消融。实时成像,温度映射和声反射光谱数据可立即评估结果。 MR对比增强图像和病理检查用于确认。结果:通过“增强”超声处理消融体积的位置是可预测的,最大可能的位移是朝向换能器6 mm,远离换能器3 mm。超声增强后的平均消融体积是常规超声相同能量下的消融体积的1.8倍。病理结果显示在增强的超声和常规超声中相同的热诱导损伤模式。结论:与标准超声相比,增强超声处理的MRgFUS是一种安全,可控且更有效的组织消融方式。这项新技术可以显着减少肿瘤消融程序的时间。 (您不是在说MRgFUS的新技术吗?如果是这样,则无需在本句末重复。)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号