Biomechanical analysis of the lumbar spine with anterior interbody fusion on the different locations of the bone grafts.

机译:腰椎前路椎间融合术在不同位置植骨的生物力学分析。

获取原文
获取原文并翻译 | 示例

摘要

The anterior lumbar interbody fusion is the common procedure in the management of the degenerated disc in the lumbar spine, but the biomechanical behavior of the fused segment would be changed because of the implantation of bone graft at the different locations. To investigate the biomechanical alteration, the study applied the finite element model to undergo the stress analysis.A three-dimensional finite element model of the lumbar spine was established, and modified to the three fusion models consisted of the bone graft at the anterior site, the middle site and the posterior site, respectively. The 12 N m flexion and the 10 N m torsion with pre-load 150 N were imposed on the L1 vertebral body.The results of the finite element model indicated that placing bone graft at anterior site could effectively resist flexion moment, and decreased the tensile force of the posterior ligaments about 15% above. Placing bone graft at posterior site could resist torsional moment, and also led to none of contact forceof the facet joint in the fused segment. However, wherever the bone grafts were placed, stress slightly increased on the disc adjacent to interbody fusion about 5% below.
机译:腰椎前路椎间融合术是处理腰椎变性椎间盘的常见方法,但由于在不同位置植入骨移植物,融合节段的生物力学行为将发生改变。为了研究生物力学变化,本研究应用有限元模型进行了应力分析。建立了腰椎的三维有限元模型,并将其修改为由前部骨移植组成的三个融合模型,中间部位和后方部位。在L1椎体上施加12 N m的屈曲力和预紧力150 N的10 N m的扭力。有限元模型的结果表明,将骨移植物置于前位可以有效抵抗屈曲力矩,并降低拉伸强度。后韧带的力量大约在15%以上。将骨移植物放置在后位点可以抵抗扭转力矩,并且也不会导致融合段中小关节的接触力。但是,无论将骨移植物放在何处,与椎间融合附近的椎间盘上的应力都会略微增加约5%以下。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号