Evaluation of factors determining the precision of body composition measurements by air displacement plethysmography.

机译:评估通过空气体积体积描记法确定人体成分测量精度的因素。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVES: To investigate methodological precision of air displacement plethysmography for assessment of body composition in a heterogenous sample of adults. DESIGN: Accuracy of volume measurements by air displacement plethysmography (ADP) for a range of known volumes was ascertained. Repeated measurements of body volume, lung volume, and derived body composition using the BODPOD measurement system were performed. Influence of surface area estimation on ADP measurement was investigated as a possible source of variation. SETTING: Clinical Nutrition Laboratory, School of Health & Sports Science, University of North London, London, UK. SUBJECTS: One hundred and two healthy subjects (57 women, 45 men) who ranged in age between 16 and 55 y and in BMI (kg/m(2)) between 17.8 and 41.9. STUDY DESIGN: Cross-sectional study of healthy adults for comparison with previous studies. Repeat measurements of raw body volume, lung volume and % body fat (BF) by ADP were all performed on the same day. RESULTS: From the range of known volumes a marked increase in the CV and a significantly greater measurement error were found at volumes below 40 l (P=0.04). Repeat measurements of raw body volume in human subjects resulted in a technical error equivalent to 0.8% BF. There was no significant difference found between measured and predicted lung volume and the 95% confidence interval for difference was only 0.3% BF. Repeat measurements of lung volume in our subset resulted in a technical error equivalent to 0.5% BF. Although body surface area estimation only accounted for variation in % BF of 0.1%, the extent of variation appeared to be governed by leanness (P<0.001). CONCLUSIONS: Although ADP retains excellent precision, in practice, repeat measurements of ADP should be performed whenever possible to allow for erroneous volume measurement within one procedure. Protocols for ADP measurement should be created with an awareness of those factors, which may affect measurements. SPONSORSHIPS: This study was supported by the University of North London Diversity & Development Fund.
机译:目的:探讨空气体积体积描记法评估异质成人样本中身体成分的方法学精度。设计:确定了通过空气体积体积描记法(ADP)对一系列已知体积进行体积测量的准确性。使用BODPOD测量系统重复测量了身体体积,肺部体积和衍生的身体成分。研究了表面积估计对ADP测量的影响,作为可能的变化来源。地点:英国伦敦北伦敦大学健康与体育科学学院临床营养实验室。受试者:一百零二名健康受试者(57名女性,45名男性)年龄在16至55岁之间,BMI(kg / m(2))在17.8至41.9之间。研究设计:健康成年人的横断面研究,与以前的研究进行比较。在同一天,通过ADP重复测量了原始体重,肺体积和体脂百分比(BF)。结果:在已知体积范围内,低于40 l的体积CV明显增加,并且测量误差明显更大(P = 0.04)。重复测量人类受试者的原始身体体积会导致技术误差等于0.8%BF。在测得的肺体积和预测的肺体积之间没有发现显着差异,差异的95%置信区间仅为0.3%BF。重复测量我们子集中的肺体积导致技术误差等于0.5%BF。尽管估计人体表面积仅占0.1%的BF的变化,但变化的程度似乎受瘦肉的支配(P <0.001)。结论:尽管ADP保留了极好的精度,但实际上,应尽可能对ADP进行重复测量,以允许在一个过程中进行错误的体积测量。应在意识到可能影响测量的那些因素的情况下创建用于ADP测量的协议。赞助:这项研究得到了北伦敦大学多样性与发展基金会的支持。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号