Delivery performance of commercial target-controlled infusion devices with Diprifusor module.

机译:带有Diprifusor模块的商业目标控制输液设备的输送性能。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND OBJECTIVE: We measured the physical performance error of commercial syringe pumps incorporating the Diprifusor module to deliver propofol by target-controlled infusion in a laboratory experiment. METHODS: Three infusion devices (Graseby 3500; Fresenius Vial Master TCI; Alaris TIVA-TCI) were studied when infusing propofol at six different target concentrations (1, 1.5, 2, 3, 5, 8 microg mL(-1)) over 10 min. This procedure was repeated five times to average each target. Precision measurements were based on the volume delivered and a mean performance error was calculated. RESULTS: The mean performance error bandwidth for all pumps was from -2.07 to 4.22% with a 95% confidence interval always <4.5%. There was a higher variability at lower target concentrations associated with lower infusion rates. The Alaris TIVA-TCI pump showed the largest overall bias whereas the Fresenius Vial Master TCI pump had the smallest divergence and the lowest variation across all target concentrations. The Graseby 3500 pump had the smallest range in mean performance error with an acceptable variation (95% CI <2%). CONCLUSIONS: Despite the differences between the pumps, which may be related to the synchronization of the hardware components, the absolute inaccuracies in the physical performance were low and presumably negligible from a clinical point of view.
机译:背景与目的:在实验室实验中,我们测量了装有Diprifusor模块通过目标控制输注输送丙泊酚的商用注射泵的物理性能误差。方法:研究了三种输注设备(Graseby 3500; Fresenius Vial Master TCI; Alaris TIVA-TCI)在六个不同的目标浓度(1、1.5、2、3、5、8 microg mL(-1))中输注丙泊酚时的浓度超过10次的情况。分钟重复该过程五次以平均每个靶标。精确测量基于所输送的体积,并计算平均性能误差。结果:所有泵的平均性能误差带宽为-2.07至4.22%,95%的置信区间始终<4.5%。在较低的目标浓度下,与较低的输注速率相关的变异性较高。 Alaris TIVA-TCI泵显示出最大的总体偏差,而Fresenius Vial Master TCI泵在所有目标浓度下的散度最小,变化最小。 Graseby 3500泵的平均性能误差范围最小,变化范围可以接受(95%CI <2%)。结论:尽管泵之间存在差异,这可能与硬件组件的同步有关,但物理性能的绝对误差很小,从临床角度来看可以忽略不计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号