首页> 外文期刊>European journal of applied physiology >VO underestimation using the Aquatrainer compared with facemask measurements: is it all in the ventilation?
【24h】

VO underestimation using the Aquatrainer compared with facemask measurements: is it all in the ventilation?

机译:与面罩测量相比,使用Aquatrainer的VO低估了:全部在通风中吗?

获取原文
获取原文并翻译 | 示例
           

摘要

After reading carefully the interesting study performed by Gay da et al. (2010), I would like to address a major concern that I have with this study. My concern has to do with the authors' interpretation of their results, in terms of the underestimation of oxygen uptake (VO_2) with the Aqua-trainer (AQ) and their subsequent conclusion that the AQ device is not suitable for use in swimming research. In the conclusions section of their paper, the authors present some possible sources of error but, in my opinion, they have neglected the possibility of a significant ventilation (V_E) measurement error.
机译:仔细阅读后,Gay da等人进行了有趣的研究。 (2010),我想解决我对这项研究的主要关注。我的担忧与作者对他们的结果的解释有关,包括对Aqua训练器(AQ)的摄氧量(VO_2)的低估以及他们随后得出的结论:该AQ设备不适合用于游泳研究。在论文的结论部分,作者提出了一些可能的误差源,但我认为他们忽略了显着的通气(V_E)测量误差的可能性。