首页> 外文期刊>Bio-medical materials and engineering >Biomechanical comparison of newly designed stemless prosthesis and conventional hip prosthesis - An experimental study.
【24h】

Biomechanical comparison of newly designed stemless prosthesis and conventional hip prosthesis - An experimental study.

机译:新设计的无茎假体和常规髋关节假体的生物力学比较-实验研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Local bone loss after implantation of traditional stem-type prostheses remains an unsolved problem during the long-term application of total hip replacement. The stress shielding effect and osteolysis were thought to be the two main factors that result in local bone loss after prosthesis implantation. A newly designed stemless cervico-trochanteric (C-T) prosthesis was thus developed to reduce stress shielding and osteolysis caused by the implantation of conventional stem-type prosthesis. Eight synthetic femora were implanted with C-T and porous coated anatomic (PCA) prostheses. Under 2,000-Newton load, the surface strains of proximal femora were compared between the intact, PCA press-fit femora and the C-T implanted femora with three different fixation modes: two-screw fixation, three-screw fixation, and three-screw combined with cement fixation. The results revealed that stress shielding in the C-T implanted femora was significantly eliminated compared to that of the PCA implanted femora (p<0.01). No statistical difference in strain magnitude was found for the C-T implanted femora among the three different fixation modes (p>0.1). The C-T implanted femur has more physiological strain distribution. Moreover, from the C-T prosthetic characteristic design, the localized osteolysis would be also reduced due to the overall coverage of neck-trochanteric area. The newly designed C-T prosthesis may be a useful alternative to the traditional stem-type prosthesis in the future.
机译:在长期使用全髋关节置换术期间,植入传统的干式假体后局部骨丢失仍未解决。应力屏蔽作用和溶骨作用被认为是造成假体植入后局部骨质流失的两个主要因素。因此,开发了一种新设计的无茎颈转子(C-T)假体,以减少因植入传统干型假体而引起的应力屏蔽和骨溶解。八个人造股骨植入C-T和多孔涂层解剖(PCA)假体。在2,000牛顿载荷下,比较完整,PCA压入配合股骨和CT植入股骨在三种不同固定方式下的股骨近端表面应变:两螺钉固定,三螺钉固定以及三螺钉结合水泥固定。结果表明,与PCA植入股骨相比,C-T植入股骨的应力屏蔽得到了显着消除(p <0.01)。在三种不同的固定方式之间,C-T植入股骨的应变幅度没有统计学差异(p> 0.1)。 C-T植入股骨具有更多的生理应变分布。此外,从C-T假体特征设计来看,由于颈粗隆区的整体覆盖范围,局部骨溶解也将减少。新设计的C-T假体可能会在将来替代传统的干式假体。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号