Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET.

机译:一种新型的完全可编程的血液采样装置的特征,用于监测PET期间的血液放射性。

获取原文
获取原文并翻译 | 示例
           

摘要

The first performance tests of a new fully programmable blood sampling device for monitoring blood radioactivity during positron emission tomography (PET) are described. Blood is withdrawn through 1-mm internal diameter tubing using an infusion pump which can be operated at rates varying from 0 to 600 ml/h. Activity in blood is measured by a 6-cm-thick bismuth germanate crystal connected to a photomultiplier tube and multichannel analyser (MCA) which are positioned within 6 cm lead shielding. Positioning of the tubing is an exact and simple procedure. The minimal readout time of the MCA is 1 s. Two independent energy windows can be set. Operation of the pump and MCA is fully controlled by a PC, i.e. sampling time, interval time and pump rate can be varied at any time during the PET scan by user-defined scripts. A number of characteristics of the new system were studied, such as sensitivity, dead time, linearity, effect of background radiation and pump rate as a function of input pressure. In addition, dispersion was measured as a function of pump rate. Finally, first clinical results were compared with manual samples. The sensitivity equalled 0.7 and 0.2 cps/Bq for 511- and 1022-keV 30% energy windows, respectively, and the system dead time was 500 ns. The system remained linear within 2% with activity concentrations up to 2.5 MBq/cc. Short-term reproducibility was better than 3% for a 1-h period. Long-term reproducibility was about 5% (ISD), which was mainly caused by variation in the diameter of the tubing. If the device was positioned in such a way that maximum shielding was directed towards the patient, the effects of background radiation from the patient on the measured activity concentration for clinically relevant conditions was minimal (<3%). Pump rate varied with input pressure, but remained constant for a given pressure. Dispersion constants smaller than 0.14 s(-1) were observed for pump rates higher than 300 ml/h, indicating that the system dispersion is small. Clinical data showed an excellent agreement to within 3% (ISD) between the results obtained with the new system and manual samples. With the continuous blood sampler radioactivity in blood can be measured accurately during the entire course of the PET scan. Furthermore, the system is fully programmable allowing adjustment of all parameters during a single PET scan.
机译:描述了一种新型的完全可编程的血液采样设备的首次性能测试,该设备用于在正电子发射断层扫描(PET)期间监测血液放射性。使用输液泵通过1毫米内径的管道抽取血液,输液泵的运行速度可在0至600毫升/小时之间变化。血液中的活性通过连接到光电倍增管和多通道分析仪(MCA)的6厘米厚的锗酸铋晶体测量,该晶体位于6厘米铅屏蔽层内。管道的定位是一个精确而简单的过程。 MCA的最小读取时间为1 s。可以设置两个独立的能量窗口。泵和MCA的操作完全由PC进行控制,即在PET扫描期间,可以通过用户定义的脚本随时更改采样时间,间隔时间和泵速。研究了新系统的许多特性,例如灵敏度,空载时间,线性,背景辐射的影响以及泵速与输入压力的关系。另外,测量分散度与泵速的关系。最后,将最初的临床结果与人工样本进行了比较。对于511-keV和1022-keV 30%能量窗口,灵敏度分别等于0.7和0.2 cps / Bq,系统死区时间为500 ns。该系统在2%范围内保持线性,活性浓度高达2.5 MBq / cc。在1小时内,短期重现性优于3%。长期可重复性约为5%(ISD),这主要是由于管道直径的变化所致。如果将设备以最大屏蔽指向患者的方式放置,则患者的背景辐射对临床相关情况下测得的活动浓度的影响极小(<3%)。泵速随输入压力而变化,但在给定压力下保持恒定。对于高于300 ml / h的泵速,观察到分散常数小于0.14 s(-1),这表明系统分散很小。临床数据显示,新系统和手动样品获得的结果之间的一致性极佳,达到3%(ISD)。使用连续采血器,可以在整个PET扫描过程中准确测量血液中的放射性。此外,该系统是完全可编程的,允许在一次PET扫描期间调整所有参数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号