PURPOSE: A joint initiative of the European Association of Nuclear Medicine (EANM) Neuroimaging Committee and EANM Research Ltd. aimed to generate a European database of [(123)I]FP-CIT single photon emission computed tomography (SPECT) scans of healthy controls. This study describes the characterization and harmonization of the imaging equipment of the institutions involved. METHODS: (123)I SPECT images of a striatal phantom filled with striatal to background ratios between 10:1 and 1:1 were acquired on all the gamma cameras with absolute ratios measured from aliquots. The images were reconstructed by a core lab using ordered subset expectation maximization (OSEM) without corrections (NC), with attenuation correction only (AC) and additional scatter and septal penetration correction (ACSC) using the triple energy window method. A quantitative parameter, the simulated specific binding ratio (sSBR), was measured using the Southampton compared against the actual values obtained from the aliquots. Camera-specific recovery coefficients were derived from linear regression and the error of the measurements was evaluated using the coefficient of variation (COV). RESULTS: The relationship between measured and actual sSBRs was linear across all systems. Variability was observed between different manufacturers and, to a lesser extent, between cameras of the same type. The NC and AC measurements were found to underestimate systematically the actual sSBRs, while the ACSC measurements resulted in recovery coefficients close to 100% for all cameras (AC range 69-89%, ACSC range 87-116%). The COV improved from 46% (NC) to 32% (AC) and to 14% (ACSC) (p < 0.001). CONCLUSION: A satisfactory linear response was observed across all cameras. Quantitative measurements depend upon the characteristics of the SPECT systems and their calibration is a necessary prerequisite for data pooling. Together with accounting for partial volume, the correction for scatter and septal penetration is essential for accurate quantification.
机译:目的:欧洲核医学协会(EANM)神经影像委员会和EANM Research Ltd.的一项联合计划,旨在建立健康对照的[(123)I] FP-CIT单光子发射计算机断层扫描(SPECT)扫描的欧洲数据库。 。这项研究描述了相关机构的成像设备的特性和协调性。方法:在所有伽马相机上获取(123)I SPECT图像,其中填充有纹状体与背景之比在10:1和1:1之间的纹状体模,并从等分试样中测量绝对比率。图像由核心实验室使用不带校正的有序子集期望最大化(OSEM)(NC),仅带衰减校正(AC)以及使用三重能量窗口方法的附加散射和间隔渗透校正(ACSC)重建。使用南安普敦测量定量参数,即模拟的特异性结合率(sSBR),与从等分试样获得的实际值进行比较。摄像机特定的恢复系数来自线性回归,并使用变异系数(COV)评估了测量误差。结果:在所有系统中,实测sSBR与实际sSBR之间的关系是线性的。观察到不同制造商之间的差异,并且在较小程度上观察到相同类型的摄像机之间的差异。发现NC和AC测量系统地低估了实际的sSBR,而ACSC测量导致所有摄像机的恢复系数接近100%(AC范围为69-89%,ACSC范围为87-116%)。 COV从46%(NC)提高到32%(AC)和14%(ACSC)(p <0.001)。结论:在所有摄像机上均观察到令人满意的线性响应。定量测量取决于SPECT系统的特性,其校准是数据合并的必要前提。除部分体积外,散射和间隔渗透的校正对于准确定量也至关重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号