首页> 外文期刊>European Journal of Nuclear Medicine and Molecular Imaging >Optimisation of protocol for low dose CT-derived attenuation correction in myocardial perfusion SPECT imaging.
【24h】

Optimisation of protocol for low dose CT-derived attenuation correction in myocardial perfusion SPECT imaging.

机译:心肌灌注SPECT成像中低剂量CT衍生衰减校正方案的优化。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: In clinical routine, attenuation correction (AC) using X-ray CT is a relatively new method for reducing attenuation artefacts. We evaluated the quality of attenuation maps generated with very low tube current to minimise exposure due to transmission scanning. METHODS: SPECT/CT acquisitions were performed with a Millenium VG3 gamma camera with the Hawkeye CT device (GE Medical Systems). In phantom studies, determination of linear absorption coefficients (mu) for air, water and Teflon was carried out. The attenuation maps in both stress and resting studies from 62 patients (21 females and 41 males, age 63.7 +/- 11.0 years, BMI 30.0 +/- 5.7 kg/m(2)) were compared. All patients underwent exercise or pharmacologic stress testing and a resting study for comparison using Tc-99m MIBI or Tc-99m Tetrofosmin. AC in stress studies was performed using 2.5 mA tube current (set as default), whereas 1.0 mA was used in resting studies. RESULTS: In both phantom and patient studies, differences of linear absorption coefficients were not significant (p > 0.05). Effective dose decreased from 0.90 mSv down to 0.36 mSv, respectively. CONCLUSION: Our results indicate that reliable attenuation maps (mu-maps) of the thorax can be obtained even with the use of very low tube current. In our study, radiation exposure in CT-based AC for myocardial perfusion SPECT was substantially lowered (60% reduction). This is of particular importance in high-risk patients who may have to undergo follow-up scans and in research studies on volunteers. The procedure introduced is relatively simple and can be transferred to other SPECT/CT devices, which allow adjustment of tube current.
机译:目的:在临床常规中,使用X射线CT进行衰减校正(AC)是减少衰减伪影的一种相对较新的方法。我们评估了非常低的管电流产生的衰减图的质量,以最大程度地减少由于透射扫描而引起的暴露。方法:使用带有Hawkeye CT装置(GE Medical Systems)的Millenium VG3伽马相机进行SPECT / CT采集。在幻像研究中,确定了空气,水和特富龙的线性吸收系数(μ)。比较了62位患者(21位女性和41位男性,年龄63.7 +/- 11.0岁,BMI 30.0 +/- 5.7 kg / m(2))在压力和静息研究中的衰减图。使用Tc-99m MIBI或Tc-99m Tetrofosmin对所有患者进行运动或药理压力测试以及静息研究以进行比较。交流电在应力研究中使用2.5 mA管电流(默认设置)进行,而静息研究中使用1.0 mA。结果:在幻像和患者研究中,线性吸收系数的差异均不显着(p> 0.05)。有效剂量分别从0.90 mSv降低至0.36 mSv。结论:我们的结果表明,即使使用非常低的管电流,也可以获得可靠的胸腔衰减图(mu-map)。在我们的研究中,基于CT的AC中心肌灌注SPECT的辐射暴露大大降低(降低了60%)。这对于可能需要进行随访扫描的高风险患者以及对志愿者的研究尤为重要。引入的过程相对简单,可以转移到其他SPECT / CT设备中,从而可以调节管电流。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号