Application of transmission scan-based attenuation compensation to scatter-corrected thallium-201 myocardial single-photon emission tomographic images.

机译:基于透射扫描的衰减补偿在散射校正th 201心肌单光子发射断层图像中的应用。

获取原文
获取原文并翻译 | 示例
       

摘要

A practical method for scatter and attenuation compensation was employed in thallium-201 myocardial single-photon emission tomography (SPET or ECT) with the triple-energy-window (TEW) technique and an iterative attenuation correction method by using a measured attenuation map. The map was reconstructed from technetium-99m transmission CT (TCT) data. A dual-headed SPET gamma camera system equipped with parallel-hole collimators was used for ECT/TCT data acquisition and a new type of external source named "sheet line source" was designed for TCT data acquisition. This sheet line source was composed of a narrow long fluoroplastic tube embedded in a rectangular acrylic board. After injection of 99mTc solution into the tube by an automatic injector, the board was attached in front of the collimator surface of one of the two detectors. After acquiring emission and transmission data separately or simultaneously, we eliminated scattered photons in the transmission and emission data with the TEW method, and reconstructed both images. Then, the effect of attenuation in the scatter-corrected ECT images was compensated with Chang's iterative method by using measured attenuation maps. Our method was validated by several phantom studies and clinical cardiac studies. The method offered improved homogeneity in distribution of myocardial activity and accurate measurements of myocardial tracer uptake. We conclude that the above correction method is feasible because a new type of 99mTc external source may not produce truncation in TCT images and is cost-effective and easy to prepare in clinical situations.
机译:al 201心肌单光子发射断层显像(SPET或ECT)中采用了一种实用的散射和衰减补偿方法,该方法采用三能窗(TEW)技术,并使用测得的衰减图进行迭代衰减校正。该地图是根据-99m传输CT(TCT)数据重建的。配备有平行孔准直仪的双头SPET伽马相机系统用于ECT / TCT数据采集,并设计了一种新型的名为“工作表线源”的外部源来进行TCT数据采集。此工作表线源由嵌入矩形丙烯酸板的细长氟塑料管组成。用自动注射器将99mTc溶液注入试管后,将板固定在两个检测器之一的准直仪表面的前面。在分别或同时获取发射和发射数据后,我们用TEW方法消除了发射和发射数据中的散射光子,并重建了两个图像。然后,通过使用测得的衰减图,用Chang的迭代方法补偿散射校正后的ECT图像中的衰减效果。我们的方法已通过多项体模研究和临床心脏研究验证。该方法改善了心肌活动的分布均匀性,并精确测量了心肌示踪剂的摄取。我们得出结论,上述校正方法是可行的,因为新型的99mTc外部源可能不会在TCT图像中产生截断,并且具有成本效益,并且在临床情况下易于制备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号