首页> 外文期刊>Europace: European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology >Non-determinable defibrillation threshold and inefficacy of implantable cardioverter/defibrillator shocks due to defective connections of the defibrillator lead terminals in the device header port.
【24h】

Non-determinable defibrillation threshold and inefficacy of implantable cardioverter/defibrillator shocks due to defective connections of the defibrillator lead terminals in the device header port.

机译:由于设备插头端口中除颤器引线端子的连接不良,导致无法确定的除颤阈值和可植入的心脏复律器/除颤器电击效率低下。

获取原文
获取原文并翻译 | 示例
           

摘要

Inefficacy of implantable cardioverter/defibrillator (ICD) shocks or inappropriate shocks in ICD patients may occur due to mechanical or technical failure of the device. In our unusual case of ICD defibrillation failure, faulty insertion and configuration of the defibrillation lead in the header port resulted in ineffective shock therapies and a non-determinable defibrillation threshold. In cases of successless defibrillations during or after ICD implantation, this problem should be considered.
机译:由于设备的机械或技术故障,可能导致植入式心脏复律器/除颤器(ICD)电击无效或ICD患者遭受不适当的电击。在我们异常的ICD除颤失败案例中,除颤导线在集管端口中的错误插入和配置导致无效的电击疗法和无法确定的除颤阈值。如果在植入ICD期间或之后进行成功的除颤,则应考虑该问题。