Troubleshooting of small R-wave with inappropriate implantable cardioverter-defibrillator shock

机译:植入式心脏复律除颤器电击不当导致小R波的故障排除

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A 5-year-old girl with catecholaminergic polymorphic ventricular tachycardia received an implantable cardioverter-defibrillator (ICD) in abdominal position after she had survived ventricular fibrillation. A shock lead was placed in the sinus transversus and a bipolar epicardial pace-sense lead was actively fixed to the left ventricle. Over time, the R-wave amplitude decreased from 10.3 mV post-operatively to finally <2 mV 2.5 years later (Panel A), resulting in T-wave oversensing and inaproppriate shock.
机译:一名患有儿茶酚胺能性多形性室性心动过速的5岁女孩在室颤后幸存下来,在腹部接受了植入式心脏复律除颤器(ICD)。将电击导线置于横窦内,并将双极心外膜心律感应导线主动固定在左心室。随着时间的流逝,R波振幅从术后的10.3 mV降低到2.5年后的<2 mV(图A),从而导致T波感应过度和不适当的电击。