Pacing electrode selection in a quadripolar left heart lead determines presence or absence of phrenic nerve stimulation.

机译:在四极左心导联中选择电极的步调可确定presence神经刺激的存在与否。

获取原文
获取原文并翻译 | 示例
           

摘要

A 60-year-old ischaemic patient presented for routine cardiac resynchronization therapy (CRT)-D implantation. An investigational quadripolar left ventricular lead was placed in the posterolateral vein. Phrenic nerve stimulation (PNS) was observed, but it occurred during pacing from only one of the four electrodes. A lead with multiple pacing electrodes is a potential alternative to physical adjustment of the lead or discontinuing CRT when PNS occurs.
机译:一名60岁的缺血患者接受了常规心脏再同步治疗(CRT)-D植入。研究性四极左心室导线放置在后外侧静脉中。观察到神经刺激(PNS),但仅在四个电极之一起搏时才发生。当出现PNS时,具有多个起搏电极的引线是对引线进行物理调整或中止CRT的潜在替代方法。