摘要

Boron-doped diamond (BDD) electrodes were used to investigate the possibility of detecting trace levels of lead by linear-sweep anodic stripping voltammetry. The low limit of detection (2 nM) is an advantage compared to other electrode materials, and it was found that at low pH values, copper concentrations that are usually present in drinking water do not affect to a large extent the detection of lead. These findings recommend anodic stripping voltammetry at the BDD electrodes as a suitable mercury-free method for the determination of trace levels of lead in drinking water. The results obtained for the lead detection in tap water real samples are in excellent agreement with those found by inductively coupled plasma-mass spectrometry (ICP-MS), demonstrating the practical analytical utility of the method.
机译:掺硼金刚石(BDD)电极用于研究通过线性扫描阳极溶出伏安法检测痕量铅的可能性。与其他电极材料相比,低检测限(2 nM)是一个优势,并且发现在低pH值下,饮用水中通常存在的铜浓度不会在很大程度上影响铅的检测。这些发现建议在BDD电极上进行阳极溶出伏安法,作为测定饮用水中痕量铅的合适无汞方法。自来水真实样品中铅检测的结果与电感耦合等离子体质谱法(ICP-MS)所发现的结果非常吻合,证明了该方法的实际分析实用性。