This work reports the construction and characterization of plastic electrochemical micro-flow-cells with integrated injection-moulded polymer electrodes. The three electrodes (working, auxiliary, and reference) were fabricated by injection-moulding from a conducting grade of polystyrene loaded with carbon fibers. On-chip reference electrodes were prepared by coating one of the conducting polymer electrodes with a Ag/AgCl layer (implemented either by e-beam evaporation of Ag followed by electrochemical formation of AgCl or by applying a Ag/AgCl paste). Working electrodes were either polymer electrodes coated with Au by e-beam evaporation or bare conducting polymer electrodes. The electrodes were integrated into the micro-flow-cells by an over-moulding process followed by ultrasonic welding. The devices were characterized by optical and electrochemical techniques. Studies by cyclic voltammetry (CV), anodic stripping voltammetry (ASV) and electrochemiluminescence (ECL) demonstrate 'proof-of-principle',of the micro-flow-cells as electrochemical sensors.
机译:这项工作报告了带有集成注塑聚合物电极的塑料电化学微流槽的构造和特性。这三个电极(工作电极,辅助电极和参比电极)是通过将导电级的装有碳纤维的聚苯乙烯注塑制成的。通过在一个导电聚合物电极上涂覆一个Ag / AgCl层来制备片上参考电极(通过电子束蒸发Ag,然后电化学形成AgCl或通过施加Ag / AgCl糊剂来实现)。工作电极可以是通过电子束蒸发涂覆了金的聚合物电极,也可以是裸露的导电聚合物电极。通过包覆成型工艺,然后进行超声波焊接,将电极集成到微流槽中。该器件通过光学和电化学技术进行了表征。通过循环伏安法(CV),阳极溶出伏安法(ASV)和电化学发光(ECL)进行的研究表明,微流通池作为电化学传感器具有“原理证明”。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号