首页> 外文期刊>Electroanalysis >Determination of mercury by chronopotentiometric stripping analysis using glassy carbon vessel as a working electrode
【24h】

Determination of mercury by chronopotentiometric stripping analysis using glassy carbon vessel as a working electrode

机译:玻碳容器作为工作电极的计时电位溶出分析法测定汞

获取原文
获取原文并翻译 | 示例

摘要

In this work chronopotentiometric stripping analysis of mercury was performed using a process vessel of glassy carbon as a working electrode. Experimental parameters for both steady and rotating vessel were investigated and compared. Relative sensitivity as well as the sharpness of the analytical signal was better when the rotation of the process vessel was performed. The new design provided better sensitivity when compared to both stationary and rotating glassy carbon disk electrodes. Detection limit obtained for the electrolysis time of 600 s was 0.1 ng/L of Hg(II). The accuracy of the technique was confirmed by analyzing the standard reference material - tomato leaves.
机译:在这项工作中,使用玻璃碳处理容器作为工作电极对汞进行了计时电位溶出分析。对稳态和旋转容器的实验参数进行了研究和比较。当进行处理容器的旋转时,相对灵敏度以及分析信号的清晰度更好。与固定式和旋转式玻璃碳圆盘电极相比,新设计提供了更好的灵敏度。电解时间为600 s的检测极限为0.1 ng / L的Hg(II)。通过分析标准参考物质-番茄叶,证实了该技术的准确性。