首页> 外文期刊>Electroanalysis >A cholesterol biosensor based on entrapment of cholesterol oxidase in a silicic sol-gel matrix at a Prussian blue modified electrode
【24h】

A cholesterol biosensor based on entrapment of cholesterol oxidase in a silicic sol-gel matrix at a Prussian blue modified electrode

机译:基于在普鲁士蓝修饰电极上的硅溶胶凝胶基质中截留胆固醇氧化酶的胆固醇生物传感器

获取原文
获取原文并翻译 | 示例
           

摘要

A cholesterol biosensors fabricated by immobilization of cholesterol oxidase (ChOx) in a layer of silicic sol-gel matrix on the top of a Prussian Blue-modified glassy carbon electrode was prepared. It is based on the detection of hydrogen peroxide produced by ChOx at - 0.05 V The half-lifetime of the biosensor is about 35 days. Cholesterol can be determined in the concentration range of 1 x 10(-6) - 8 x 10(-5) mol/L with a detection limit of 1.2 x 10(-7) mol/L. Normal interfering compounds, such as ascorbic acid and uric acid do not affect the determination. The high sensitivity and outstanding selectivity are attributed to the Prussian Blue film modified on the sensor. [References: 41]
机译:制备了胆固醇生物传感器,该传感器通过将胆固醇氧化酶(ChOx)固定在普鲁士蓝修饰的玻碳电极顶部的硅溶胶凝胶基质中制成。它基于检测ChOx在-0.05 V时产生的过氧化氢。生物传感器的半衰期约为35天。胆固醇的浓度范围为1 x 10(-6)-8 x 10(-5)mol / L,检测极限为1.2 x 10(-7)mol / L。正常的干扰化合物,例如抗坏血酸和尿酸不会影响测定。高灵敏度和出色的选择性归因于传感器上修饰的普鲁士蓝膜。 [参考:41]