On-column preconcentration and separation of DNA fragments using polymer solutions in the presence of electroosmotic flow

机译:在电渗流存在下使用聚合物溶液进行柱上预浓缩和DNA片段分离

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We demonstrated DNA preconcentration and separation in the presence of electroosmotic flow (EOF) using poly(ethylene oxide) (PEO) solutions. After injecting large volumes of DNA samples into a capillary filled with free tris(hydroxymethyl)aminomethane (Tris)-borate (TB) buffers, PEO solutions entered the capillary by EOF and acted as sieving matrices. In contrast to conventional methods (in the absence of EOF), controlling the EOF was also useful for resolution optimization. We have found that PEO adsorption on the capillary wall was more pronounced when low ionic strength buffers were used. Thus, the EOF decreased with increasing injection length, which led to longer migration times and changes in resolution and stacking efficiency. All resolution values were higher than 1.5 when 1.0 mu g/mL DNA samples were injected at 240 V/cm for 60 s (0.67 mu L). In addition, as low as 0.015 mu g/mL DNA samples (an about 66-fold increase in sensitivity) were detected when the injection was performed at 250 V/cm for 60 s. [References: 31]
机译:我们展示了在使用聚环氧乙烷(PEO)溶液的电渗流(EOF)存在下的DNA预浓缩和分离。将大量的DNA样品注入装有游离三(羟甲基)氨基甲烷(Tris)-硼酸盐(TB)缓冲液的毛细管中后,PEO溶液通过EOF进入毛细管并用作筛分基质。与常规方法(在没有EOF的情况下)相比,控制EOF对于分辨率优化也很有用。我们发现,当使用低离子强度缓冲液时,PEO在毛细管壁上的吸附更为明显。因此,EOF随着注射长度的增加而降低,这导致更长的迁移时间以及分辨率和堆叠效率的变化。当以240 V / cm注入1.0μg / mL DNA样品60 s(0.67μL)时,所有分辨率值均高于1.5。此外,当以250 V / cm的速度进样60 s时,检测到的DNA样品低至0.015μg / mL(灵敏度提高了约66倍)。 [参考:31]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号