A microfluidic device with integrated waveguides and a long path length detection cell for UV/Vis absorbance detection is presented. The 750 mum U-cell detection geometry was evaluated in terms of its optical performance as well as its influence on efficiency for electrophoretic separations in the microdevice. Stray light was found to have a strong effect on both, the sensitivity of the detection and the available linear range. The long path length U-cell showed a 9 times higher sensitivity when compared to a conventional capillary electrophoresis (CE) system with a 75 mum inner diameter (ID) capillary, and a 22 times higher sensitivity than with a 50 mum ID capillary. The linear range was comparable to that achieved in a 75 mum ID capillary and more than twice as large as in a 50 mum ID capillary. The use of the 750 mum U-cell did not contribute significantly to band broadening; however, a clear quantification was made difficult by the convolution of several other band broadening sources. [References: 20]
机译:提出了一种具有集成波导和用于UV / Vis吸收检测的长程检测池的微流体装置。根据其光学性能及其对微器件中电泳分离效率的影响,评估了750微米的U细胞检测几何形状。发现杂散光对检测的灵敏度和可用的线性范围都有很强的影响。与具有75微米内径(ID)毛细管的常规毛细管电泳(CE)系统相比,长路径长度的U池显示出9倍的灵敏度,而与50微米ID的毛细管相比,其灵敏度高22倍。线性范围可与内径为75微米的毛细管相比,是内径为50微米的毛细管的两倍以上。 750毫米的U细胞的使用并没有显着促进频带的扩展。但是,由于其他几种频带扩展源的卷积,很难进行清晰的量化。 [参考:20]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号