A new capillary electrophoresis end-column amperometric detection system without the need for capillary/electrode alignment

机译:无需毛细管/电极对准的新型毛细管电泳末端柱安培检测系统

获取原文
获取原文并翻译 | 示例

摘要

A self-aligning end-column amperometric detection system for capillary electrophoresis was constructed. In this system, the electrode and capillary were exchanged easily and the capillary/electrode alignment procedure is not required. Gold, gold/mercury amalgam, copper and carbon fiber could be used as the working electrode. The principle is in the use of two disk holders with the capillary and the electrode in the center, so that by inserting the disk holders into a groove in the working electrode port, the capillary and the electrode are automatically aligned and the distance between the capillary and the electrode is assured at 0.24 mm. The relative standard deviation obtained using five different gold/mercury amalgam microdisk electrodes for determination of cysteine was 1.5% for the migration time and 3.3% for the electrophoretic peak current. The simple and convenient system was attractive for the routine analysis by capillary electrophoresis with electrochemical detection. The system was applied to the determination of promethazine hydrochloride in human serum. [References: 36]
机译:建立了用于毛细管电泳的自对准末端柱安培检测系统。在该系统中,电极和毛细管易于更换,并且不需要毛细管/电极对齐程序。金,金/汞合金,铜和碳纤维可用作工作电极。原理是使用两个圆盘固定器,毛细管和电极位于中心,因此,通过将圆盘固定器插入工作电极端口的凹槽中,毛细管和电极将自动对齐,并且毛细管之间的距离并确保电极为0.24 mm。使用五个不同的金/汞齐汞齐微盘电极测定半胱氨酸获得的相对标准偏差,其迁移时间为1.5%,电泳峰电流为3.3%。这种简单方便的系统对于通过毛细管电泳和电化学检测进行常规分析很有吸引力。该系统用于人血清中盐酸异丙嗪的测定。 [参考:36]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号