Trace determination of perchlorate using electromembrane extraction and capillary electrophoresis with capacitively coupled contactless conductivity detection.

机译:使用电膜萃取和毛细管电泳与电容耦合非接触电导检测进行痕量高氯酸盐测定。

获取原文
获取原文并翻译 | 示例
           

摘要

Electromembrane extraction (EME) and CE with capacitively coupled contactless conductivity detection (CE-C(4) D) was applied to rapid and sensitive determination of perchlorate in drinking water and environmental samples. Porous polypropylene hollow fiber impregnated with 1-heptanol acted as a supported liquid membrane (SLM) and perchlorate was transported and preconcentrated in the fiber lumen on application of electric field. High selectivity of perchlorate determination and its baseline separation from major inorganic anions was achieved in CE-C(4) D using background electrolyte solution consisting of 7.5 mM L-histidine and 40 mM acetic acid at pH 4.1. The analytical method showed excellent parameters in terms of reproducibility; RSD values for migration times and peak areas at a spiked concentration of 15 mug/L of perchlorate (US EPA recommended limit for drinking water) were below 0.2 and 8.7%, respectively, in all examined water samples. Linear calibration curves were obtained for perchlorate in the concentration range 1-100 mug/L (r(2) >/=0.999) with limits of detection at 1 mug/L for tap water and at 0.25-0.35 mug/L for environmental and bottled potable water samples. Recoveries at 15 mug/L of perchlorate were between 95.9 and 106.7% with minimum and maximum recovery values for snow and bottled potable water samples, respectively.
机译:电膜萃取(EME)和带有电容耦合的非接触电导率检测(CE-C(4)D)的CE用于饮用水和环境样品中高氯酸盐的快速灵敏测定。用1-庚醇浸渍的聚丙烯中空多孔纤维充当支撑液膜(SLM),在施加电场的情况下,高氯酸盐在纤维管腔中运输和预浓缩。使用背景电解质溶液(由7.5 mM L-组氨酸和40 mM乙酸组成,pH 4.1),在CE-C(4)D中实现了高氯酸根测定及其与主要无机阴离子的基线分离的高选择性。该分析方法在重现性方面显示出极好的参数;在所有检查的水样品中,高氯酸盐的峰值浓度为15杯/升(美国EPA建议的饮用水限值)时,迁移时间和峰面积的RSD值分别低于0.2和8.7%。对于浓度范围为1-100杯/ L(r(2)> / = 0.999)的高氯酸盐,获得了线性校准曲线,自来水的检测限为1杯/ L,环境和环境的检测限为0.25-0.35杯/ L。瓶装饮用水样品。 15杯/升高氯酸盐的回收率在95.9至106.7%之间,雪和瓶装饮用水样品的最低和最高回收率分别为。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号