Ready-to-use screen printed glucose sensors are fabricated using Prussian Blue (PB) and Cobalt Phthalocyanine (CoPC) mediated carbon inks as working electrodes. The reference and counter electrodes are screen printed using silver/silver chloride and graphitic carbon paste respectively. The screen printed reference electrodes (internal reference electrode (IRE)) are found to be stable for more than 60 minutes when examined with saturated calomel electrode. Optimal operating voltage for PB and CoPC screen printed sensors are determined by hydrodynamic voltammetric technique. Glucose oxidase is immobilized on the working electrodes by cross-linking method. PB mediated glucose sensor exhibits a sensitivity of 5.60 mA cm(-2)/mM for the range, 10 to 1000 mu M. Sensitivity of CoPC mediated glucose sensor is found to be 5.224 mu A cm(-2)/mM and amperometeric response is linear for the range, 100 to 1500 mu M. Interference studies on the fabricated glucose sensors are conducted with species like uric acid and ascorbic acid. PB mediated sensors showed a completely interference-free behavior. The sensing characteristics of PB mediated glucose sensors are also studied in diluted human serum samples and the results are compared with the values obtained through standard clinical method. The co-efficient of variation is found to be less than 5%. (C) 2015 The Electrochemical Society. All rights reserved.
机译:现成的丝网印刷葡萄糖传感器使用普鲁士蓝(PB)和酞菁钴(CoPC)介导的碳墨水作为工作电极来制造。参比电极和对电极分别使用银/氯化银和石墨碳浆料进行丝网印刷。当用饱和甘汞电极检查时,发现丝网印刷的参比电极(内部参比电极(IRE))稳定60分钟以上。 PB和CoPC丝网印刷传感器的最佳工作电压通过流体动力学伏安法确定。葡萄糖氧化酶通过交联法固定在工作电极上。 PB介导的葡萄糖传感器在10至1000μM的范围内显示出5.60 mA cm(-2)/ mM的灵敏度。发现CoPC介导的葡萄糖传感器的灵敏度为5.224μA cm(-2)/ mM和安培响应在100至1500μM的范围内呈线性关系。对制造的葡萄糖传感器的干扰研究是通过尿酸和抗坏血酸等物质进行的。 PB介导的传感器显示出完全无干扰的行为。在稀释的人血清样品中还研究了PB介导的葡萄糖传感器的传感特性,并将结果与​​通过标准临床方法获得的值进行了比较。发现变化系数小于5%。 (C)2015年电化学学会。版权所有。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号