Evaluation of treatment planning system monitor unit calculations for three intensity modulated radiotherapy delivery techniques

机译:评估三种强度调制放射治疗技术的治疗计划系统监控器单元的计算

获取原文
获取原文并翻译 | 示例

摘要

Background: We have validated the monitor unit calculations from a commercially available treatment planning system (TPS) for three intensity modulated radiotherapy (IMRT) planning techniques for tangential breast irradiation by using ionization chamber measurements. Materials and Methods: Treatment plans were generated for forty-two breast patients by a forward planned field in field technique, electronic tissue compensation (ETC), and an inverse planned sliding window technique. We also performed a reproducibility of delivery and dose linearity analysis for each technique. The treatments were delivered to a phantom using a Varian CL21EX linear accelerator. A 2571 0.6 cm~3 Farmer type ionization chamber and Farmer 2570/1 electrometer from NE Technology was used to measure output of the linear accelerator and the dose at predefined point in the verification plan. Results: The agreement between the measured and calculated dose was -0.87% +- 0.54% for field in field technique, -0.74% +-0.23% for electronic tissue compensators, and -1.26% +- 0.48% for the inverse planning technique and. In terms of reproducibility the mean deviation was -1.10% +-0.44% for the field in field technique, -0.38% +- 0.42% for electronic tissue compensators, -1.04% + 0.42% for inverse planning technique. Dose linearity experiments showed no significant variations for clinical situations but a breakdown was observed in relative dose for very low monitor units. Conclusion: We have found that the monitor unit calculations for all three planning techniques are correct to the order of 1%, and that the plans can be delivered in a reproducible and accurate manner. Iran. J. Radiat. Res., 2011; 9(3): 145-150
机译:背景:我们已经通过电离室测量验证了用于切线乳房照射的三种强度调制放射疗法(IMRT)计划技术的商用治疗计划系统(TPS)的监护仪单位计算。材料和方法:通过向前计划的现场技术,电子组织补偿(ETC)和反向计划的滑动窗口技术,为42名乳腺患者制定了治疗计划。我们还对每种技术进行了传递和剂量线性分析的可再现性。使用Varian CL21EX线性加速器将治疗剂传递到体模中。 NE Technology的2571 0.6 cm〜3 Farmer型电离室和Farmer 2570/1静电计用于验证验证计划中线性加速器的输出和预定点的剂量。结果:实地技术的实测剂量与计算剂量之间的一致性为-0.87%+-0.54%,电子组织补偿器为-0.74%+ -0.23%,逆向计划技术为-1.26%+-0.48% 。就重现性而言,现场技术的平均偏差为-1.10%+ -0.44%,电子组织补偿器的平均偏差为-0.38%+-0.42%,逆向计划技术的平均值偏差为-1.04%+ 0.42%。剂量线性实验显示临床情况无明显变化,但对于非常低的监测单位,在相对剂量下观察到分解。结论:我们发现,所有三种计划技术的监视单位计算都正确到1%的数量级,并且计划可以以可重复且准确的方式交付。伊朗。 J. Radiat。 Res。,2011; 9(3):145-150

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号