Vascular enhancement in early dynamic liver MR imaging in an animal model: comparison of two injection regimen and two different doses Gd-EOB-DTPA (gadoxetic acid) with standard Gd-DTPA.

机译:在动物模型中早期动态肝脏MR成像中的血管增强:两种注射方案和两种不同剂量的Gd-EOB-DTPA(葡萄糖酸)与标准Gd-DTPA的比较。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVES: To investigate the influence of 2 different doses and injection rates on the enhancement of liver vasculature in Gd-EOB-DTPA enhanced liver MRI compared with the standard dose Gd-DTPA. MATERIALS AND METHODS: This animal experimental study has been approved by the local authorities. In 5 pigs, a time-resolved T1w 3D GRE sequence, and in a second experiment, a single-slice turbo FLASH T1w sequence with 3 frames/s were started each with contrast agent application of Gd-EOB-DTPA and Gd-DTPA. Three sets with different doses were performed with an injection rate of 1 and 2 mL/s which are as follows: A, Standard dose: 25 [mu]mol Gd-EOB-DTPA/kg body-weight (bw); B, Double dose: 50 [mu]mol Gd-EOB-DTPA/kg bw; and C, Standard dose: 100 [mu]mol Gd-DTPA/kg bw. Signal intensities were measured in aorta, vena cava inferior, portal vein, and liver parenchyma, and time-signal-to-noise (SNR)-curves were generated. The increase in SNR from baseline and several perfusion parameters were calculated. Statistical significance was tested with the Wilcoxon rank test at a significance level of P = 0.05. RESULTS: Mean peak enhancement (SNR) in the aorta was not significantly different for set A, B, and C with the same injection rate. Mean peak enhancement in the aorta was significantly higher for set A at 1 mL/s than at 2 mL/s (159.1 vs. 144.4, P = 0.043). Mean peak enhancement in the PV, the vena cava inferior, and the liver parenchyma was significantly lower for set A compared with set B and set C, at both injection speeds. The full width at half max was significantly longer for injection protocols with 1 as compared with 2 mL/s for sets A and set B. Set A with 1 mL/s reached similar values as compared with set C with 2 mL/s for the full width at half max (8.92 vs. 9.40; P = 0.35), for the area-under-the curve (1736.64 vs. 1912.74; P = 0.69) and the maximal signal-to-noise ratio (25.21 vs. 25.37; P = 0.69). CONCLUSION: Despite the lower amount of gadolinium in the standard dose of Gd-EOB-DTPA, the results showed that the arterial enhancement in Gd-EOB-DTPA-enhanced dynamic liver MRI was comparable to Gd-DTPA. This result can be explained mainly by the higher relaxivity. Choosing a lower injection rate additionally supported to compensate for the lower injection volume by stretching the bolus without decreasing the peak. In this respect, an injection rate of 1 mL/s showed better results with regard to the arterial enhancement compared with 2 mL/s.
机译:目的:研究与标准剂量Gd-DTPA相比,两种不同剂量和注射速率对Gd-EOB-DTPA增强的肝脏MRI中肝血管系统增强的影响。材料与方法:该动物实验研究已获得地方当局的批准。在5头猪中,一个时间分辨的T1w 3D GRE序列,在第二个实验中,开始分别使用Gd-EOB-DTPA和Gd-DTPA的造影剂,以3帧/秒的速度启动单片涡轮FLASH T1w序列。三组不同剂量的注射速率分别为1和2 mL / s,如下所示:A,标准剂量:25μmolGd-EOB-DTPA / kg体重(bw); B,两次剂量:50μmolGd-EOB-DTPA / kg bw;和C,标准剂量:100μmolGd-DTPA / kg bw。测量主动脉,下腔静脉,门静脉和肝实质中的信号强度,并生成时间信噪比(SNR)曲线。计算出从基线到SNR的增加以及一些灌注参数。使用Wilcoxon等级检验以P = 0.05的显着性水平检验统计学显着性。结果:在相同的注射速率下,A,B和C组主动脉的平均峰值增强(SNR)没有显着差异。组A在1 mL / s时的主动脉平均峰值增强显着高于2 mL / s(159.1对144.4,P = 0.043)。在两种注射速度下,与B组和C组相比,A组的PV,下腔静脉和肝实质的平均峰值增强显着较低。与A组和B组的2 mL / s相比,注射方案为1的半峰全宽明显更长。与A组和B组的2 mL / s的C组相比,A组的1 mL / s达到了相似的值。对于曲线下方区域(1736.64对1912.74; P = 0.69)和最大信噪比(25.21对25.37; P),最大一半时的全宽(8.92对9.40; P = 0.35) = 0.69)。结论:尽管标准剂量的Gd-EOB-DTPA中of的含量较低,但结果表明,Gd-EOB-DTPA增强的动态肝脏MRI的动脉增强作用与Gd-DTPA相当。该结果主要可以由较高的松弛度来解释。还可以选择较低的注入速率,以在不降低峰的情况下通过扩展推注来补偿较低的注入量。在这方面,与2 mL / s相比,1 mL / s的注射速率显示出更好的动脉增强效果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号