Feasibility of real-time magnetic resonance-guided angioplasty and stenting of renal arteries in vitro and in Swine, using a new polyetheretherketone-based magnetic resonance-compatible guidewire.

机译:使用新的基于聚醚醚酮的磁共振兼容导丝,在体外和在猪中进行实时磁共振引导的血管成形术和肾动脉支架置入术的可行性。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: Demonstrate the usability of a new polyetheretherketone (PEEK)-based MR-compatible guidewire for renal artery catheterization, angioplasty, and stenting under MR-guidance using MR-visible markers, in vitro and in vivo. MATERIAL AND METHODS: The new 0.035'' guidewire with fiber-reinforced PEEK core, a soft tip, and a hydrophilic coating was used. Paramagnetic markings were coated on the wire and nonbraided catheters for passive visualization. Bending stiffness of the guidewire was compared with available hydrophilic guidewires (Terumo Glidewire Stiff and Standard). A human aortic silicon phantom and 2 pigs were used. The study was animal care and use approved by the committee. Under MR-guidance, renal arteries were catheterized, balloon angioplasty was performed, and balloon expandable renal artery stents were deployed in vivo. Post mortem autopsy was performed. Guidewire visibility, pushability, steerability, and device-support capabilities of the marked guidewire were qualitatively assessed. Procedure times were recorded. RESULTS: Bending stiffness of the new PEEK-based wire was comparable with Standard Glidewire. In vitro and in vivo guidewire guidance, catheter configuration, renal artery catheterization, and balloon angioplasty were successful. In pigs, stent deployments were successful in both renal arteries. Autopsy revealed acceptable stent positioning. Guidewire visibility through applied markers was acceptable. Steerability, pushability, and device support were good in vitro and in vivo. CONCLUSIONS: The PEEK-based guide allows percutaneous MR-guided renal artery angioplasty and stenting with sufficient visibility, good steerability, pushability, and device support.
机译:目的:演示一种新的基于聚醚醚酮(PEEK)的MR兼容导丝在体外和体内使用MR可见标记在MR引导下用于肾动脉导管插入,血管成形术和支架置入的可用性。材料和方法:使用了新的0.035''导丝,该导丝具有纤维增强的PEEK芯,柔软的尖端和亲水涂层。顺磁性标记涂在导线和非编织导管上,以进行被动观察。将导丝的弯曲刚度与可用的亲水性导丝(Terumo Glidewire Stiff和Standard)进行了比较。使用了人主动脉硅体模和2头猪。该研究是委员会批准的动物护理和使用方法。在MR的指导下,对肾动脉进行导尿,进行球囊血管成形术,并在体内部署球囊扩张性肾动脉支架。进行尸检。定性评估了标记导丝的导丝可见性,可推动性,可操纵性和设备支持功能。记录手术时间。结果:新的PEEK基焊丝的弯曲刚度可与Standard Glidewire媲美。体外和体内导丝引导,导管配置,肾动脉导管插入术和球囊血管成形术均获得成功。在猪中,两个肾动脉的支架部署均成功。尸检显示可接受的支架定位。通过应用标记的导线可见性是可以接受的。在体外和体内,可操纵性,推动性和设备支持均良好。结论:基于PEEK的导向器允许经皮MR引导的肾动脉血管成形术和支架具有足够的可见度,良好的可操纵性,可推动性和设备支持。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号