Hippocampal MRI volumetry at 3 Tesla: reliability and practical guidance.

机译:3 Tesla的海马MRI容量测定:可靠性和实用指南。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVES: Although volumetry of the hippocampus is considered to be an established technique, protocols reported in literature are not described in great detail. This article provides a complete and detailed protocol for hippocampal volumetry applicable to T1-weighted magnetic resonance (MR) images acquired at 3 Tesla, which has become the standard for structural brain research. MATERIALS AND METHODS: The protocol encompasses T1-weighted image acquisition at 3 Tesla, anatomic guidelines for manual hippocampus delineation, requirements of delineation software, reliability measures, and criteria to assess and ensure sufficient reliability. Moreover, the validity of the correction for total intracranial volume size was critically assessed. The protocol was applied by 2 readers to the MR images of 36 patients with cryptogenic localization-related epilepsy, 4 patients with unilateral hippocampal sclerosis, and 20 healthy control subjects. RESULTS: The uncorrected hippocampal volumes were 2923 +/- 500 mm3 (mean +/- SD) (left) and 3120 +/- 416 mm3 (right) for the patient group and 3185 +/- 411 mm3 (left) and 3302 +/- 411 mm3 (right) for the healthy control group. The volume of the 4 pathologic hippocampi of the patients with unilateral hippocampal sclerosis was 2980 +/- 422 mm3. The inter-reader reliability values were determined: intraclass-correlation-coefficient (ICC) = 0.87 (left) and 0.86 (right), percentage volume difference (VD) = 7.0 +/- 4.7% (left) and 6.0 +/- 3.8% (right), and overlap ratio (OR) = 0.82 +/- 0.04 (left) and 0.82 +/- 0.03 (right). The positive Pearson correlation between hippocampal volume and total intracranial volume was found to be low: r = 0.48 (P = 0.03, left) and r = 0.62 (P = 0.004, right) and did not significantly reduce the volumetric variances, showing the limited benefit of the brain size correction. CONCLUSIONS: A protocol was described to determine hippocampal volumes based on 3 Tesla MR images with high inter-reader reliability. Although the reliability of hippocampal volumetry at 3 Tesla was similar to the literature values obtained at 1.5 Tesla, hippocampal border definition is argued to be more confident and easier because of the improved signal-to-noise characteristics.
机译:目的:尽管海马体量被认为是一项既定技术,但文献中报道的方案并未得到详细描述。本文提供了适用于在3特斯拉获取的T1加权磁共振(MR)图像的海马体积测量的完整而详细的协议,该协议已成为大脑结构研究的标准。材料与方法:该协议包括在3 Tesla处进行T1加权图像采集,海马体手动划定的解剖学指南,划界软件的要求,可靠性措施以及评估和确保足够可靠性的标准。而且,严格评估了颅内总体积校正的有效性。该协议由2位读者应用于36例隐源性定位相关性癫痫患者,4例单侧海马硬化患者和20例健康对照者的MR图像。结果:未经校正的海马体积分别为2923 +/- 500 mm3(平均+/- SD)(左)和3120 +/- 416 mm3(右),患者组为3185 +/- 411 mm3(左)和3302 + /-411立方毫米(右),用于健康对照组。单侧海马硬化患者的4个病理性海马的体积为2980 +/- 422 mm3。确定阅读器之间的可靠性值:类内相关系数(ICC)= 0.87(左)和0.86(右),体积差异百分比(VD)= 7.0 +/- 4.7%(左)和6.0 +/- 3.8 %(右),重叠率(OR)= 0.82 +/- 0.04(左)和0.82 +/- 0.03(右)。发现海马体积与颅内总体积之间的正Pearson相关性很低:r = 0.48(P = 0.03,左)和r = 0.62(P = 0.004,右),并且没有显着减小体积变化,显示出局限性大脑大小校正的好处。结论:描述了一种协议,该协议可基于3台Tesla阅读器MR图像确定海马体积,并具有较高的阅读器间可靠性。尽管在3特斯拉时海马容积测量的可靠性与在1.5特斯拉时获得的文献值相似,但是由于改善了信噪比特性,因此海马边界定义被认为更加自信和容易。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号