OBJECTIVE: The objective of this study was to implement a three-dimensional (3-D) T1 mapping sequence (3DLL) at 3.0 T for dGEMRIC based on the Look Locker scheme. MATERIALS AND METHODS: Because all current reports on dGEMRIC are at 1.5 T and mostly using 2-D IR fast spin echo (FSE), data were acquired at 1.5 T and 3.0 T with both 3DLL and 2-D IR-FSE sequence. Phantoms with different concentrations of Gd(DTPA) were used and seven subjects (three asymptomatic, four symptomatic) were scanned using the dGEMRIC technique. RESULTS: The T1 measurements obtained on the phantom with 3DLL show very good agreement with those acquired with 2-D IR-FSE. Using a two-tailed paired t test, the T1 (Gd) measurements in two sections obtained in all subjects with both sequences were found to be statistically indistinguishable at either field strength (P = 0.07 at 1.5 T and P = 0.07 at 3.0 T). CONCLUSIONS: The preliminary data presented here suggest that the 3DLL sequence provides accurate T1 values with sufficient in-planeresolution and allows full joint coverage in less than 10 minutes.
机译:目的:本研究的目的是基于Look Locker方案在dGEMRIC上以3.0 T实现三维(3-D)T1映射序列(3DLL)。材料与方法:由于dGEMRIC上的所有当前报告均为1.5 T,且大多使用2-D IR快速自旋回波(FSE),因此使用3DLL和2-D IR-FSE序列在1.5 T和3.0 T时采集了数据。使用具有不同浓度的Gd(DTPA)的幻影,并使用dGEMRIC技术扫描了7名受试者(3名无症状,4名有症状)。结果:用3DLL在幻象上获得的T1测量值与使用2-D IR-FSE获得的测量值非常吻合。使用两尾配对t检验,发现在所有受试者中获得两个序列的两个部分的T1(Gd)测量值在任一场强下均无统计学差异(1.5 T时P = 0.07,3.0 T时P = 0.07) 。结论:此处提供的初步数据表明3DLL序列可提供准确的T1值以及足够的面内分辨率,并允许在不到10分钟的时间内完全覆盖关节。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号