首页> 外文期刊>Investigative ophthalmology & visual science >Repeatability of manual subfoveal choroidal thickness measurements in healthy subjects using the technique of enhanced depth imaging optical coherence tomography.
【24h】

Repeatability of manual subfoveal choroidal thickness measurements in healthy subjects using the technique of enhanced depth imaging optical coherence tomography.

机译:使用增强深度成像光学相干断层扫描技术,在健康受试者中进行人工小凹下脉络膜厚度测量的可重复性。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The aim of this study was to investigate the repeatability of manual measurements of choroidal thickness in healthy subjects imaged on spectral domain optical coherence tomography (OCT) using the enhanced depth imaging (EDI) technique. METHODS: Fifty consecutive, healthy, young, adult volunteers with no known eye disease were enrolled prospectively. Two good-quality horizontal and vertical line scans through the fovea were obtained for each eye. Using the manual calipers provided by the software of the proprietary device, two experienced OCT readers measured the subfoveal choroidal thickness (SFCT) of the horizontal and vertical line scans for all eyes. The readers were masked to each other's readings. Intraobserver, interobserver, and intrasession coefficients of repeatability (CRs) were calculated. RESULTS: Mean (standard deviation [SD]) age of the study subjects was 38 (5) years (range, 30-49 years). Mean (SD) subfoveal choroidal thickness was 332 (90) mum (right eyes) and 332 (91) mum (left eyes). Intraobserver CR was approximately 23 (95% confidence interval [CI], 19-26) mum, whereas interobserver and intrasession CRs were greater at 32 (95% CI, 30-34) and 34 (95% CI, 32-36) mum, respectively. There was no significant difference in SFCT between all pairs of SFCT measurements except for the two intrasession vertical line scans. CONCLUSION: A change of >32 mum was likely to exceed interobserver variability in SFCT. Future studies are required to estimate the repeatability of SFCT measurements in patients with chorioretinal pathology.
机译:目的:本研究的目的是研究使用增强深度成像(EDI)技术在光谱域光学相干断层扫描(OCT)上成像的健康受试者中脉络膜厚度的手动测量的可重复性。方法:前瞻性招募了五十名连续,健康,年轻,成人,无已知眼病的志愿者。对于每只眼睛,均获得了两次通过中央凹的高质量水平和垂直线扫描。使用专有设备软件提供的手动卡尺,两个经验丰富的OCT读取器测量了所有眼睛的水平和垂直线扫描的中央凹脉络膜厚度(SFCT)。读者们互相掩盖了自己的读物。计算了观察者内,观察者间和会话间重复性系数(CR)。结果:研究对象的平均年龄(标准偏差[SD])为38(5)岁(范围为30-49岁)。平均(SD)凹下脉络膜厚度为332(90)妈妈(右眼)和332(91)妈妈(左眼)。观察者内CR约为23(95%置信区间[CI],19-26)妈妈,而观察者间和会话内CR则更大,分别为32(95%CI,30-34)和34(95%CI,32-36)妈妈。 , 分别。除两次会话内垂直线扫描外,所有SFCT测量对之间的SFCT均无显着差异。结论:> 32毫米的变化可能会超过SFCT的观察者间差异。需要进行进一步的研究来评估脉络膜视网膜疾病患者中SFCT测量的可重复性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号