首页> 外文期刊>Investigative ophthalmology & visual science >Repeatability of Spectralis OCT measurements of macular thickness and volume in diabetic macular edema
【24h】

Repeatability of Spectralis OCT measurements of macular thickness and volume in diabetic macular edema

机译:Spectralis OCT测量在糖尿病性黄斑水肿中黄斑厚度和体积的可重复性

获取原文
获取原文并翻译 | 示例

摘要

Purpose. We estimated coefficients of repeatability for Spectralis optical coherence tomography (OCT)-derived automated retinal thickness and volume measurements in subjects with center-involving diabetic macular edema (DME). Methods. A total of 50 eyes of 50 consecutive patients with center-involving DME underwent four consecutive "fast" volume scans at a single session using one OCT device operated by one of two experienced operators. Bland-Altman coefficients of repeatability (CR) were calculated for automated retinal thickness measurements in the nine Early Treatment of Diabetic Retinopathy Study (ETDRS) subfields, center point thickness, and total macular volume. Scans were evaluated for significant automated retinal boundary detection error and revised estimates for CR calculated with these scans excluded. Results. CR in the central subfield was 8.03 μm (95% confidence interval [CI] 7.70-8.35 μm). In other subfields, CR ranged from 6.54 to 18.25 μm. Scan sets from 13 subjects had significant boundary detection error; reanalysis with these excluded yielded a CR for the central subfield of 7.44 μm with CR for all other subfields <8 μm. Conclusions. Retinal thickness measurements in subjects with DME obtained using Spectralis OCT are considerably less variable than has been reported with other devices. Changes in central subfield thickness >8 μm can be considered more indicative of true clinical change rather than measurement variability. This finding informs clinical practice and clinical trial design.
机译:目的。我们估算了患有中枢性糖尿病性黄斑水肿(DME)的受试者的Spectralis光学相干断层扫描(OCT)衍生的自动视网膜厚度和体积测量的可重复性系数。方法。使用两名经验丰富的操作员之一操作的一台OCT设备,在单个疗程中对50位连续的中心性DME患者进行了总共50只眼睛的连续四次“快速”体积扫描。在9个糖尿病性视网膜病变早期治疗研究(ETDRS)子字段,中心点厚度和总黄斑体积中,针对自动视网膜厚度测量计算了Bland-Altman重复性系数(CR)。对扫描进行评估是否存在明显的自动视网膜边界检测错误,并排除这些扫描而计算出的CR修正值。结果。中央子场中的CR为8.03μm(95%置信区间[CI] 7.70-8.35μm)。在其他子区域中,CR范围为6.54至18.25μm。来自13位受试者的扫描集具有明显的边界检测错误;排除这些因素进行的重新分析得出中央子域的CR为7.44μm,而所有其他子域的CR均小于8μm。结论与其他设备相比,使用Spectralis OCT获得的DME患者的视网膜厚度测量的可变性要小得多。中心子区域厚度> 8μm的变化可被认为更能说明临床上的真实变化,而不是测量变异性。这一发现有助于临床实践和临床试验设计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号