Repeatability and reproducibility of macular thickness measurements with the Humphrey OCT system.

机译:使用汉弗莱OCT系统进行黄斑厚度测量的可重复性和可重复性。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: To assess the accuracy, precision, repeatability, and reproducibility of measurements made by the Humphrey optical coherence tomography (OCT) system (Humphrey-Zeiss Medical Systems, San Leandro, CA). METHODS: The performance of the system was first investigated by scanning a test object containing an air gap of known size. Measurements were repeated with water or glycerin in the gap. In the clinical setting, macular thickness measurements were obtained from a control group of 20 normal subjects. For analysis, these scans were divided into eight sections, each containing 10 A-scans. RESULTS: The average gap thickness was found to be close to the true value in all cases. The overall coefficients of intersession reproducibility were less than 1% for the test object and 1.51% for the control group. There was no significant difference between scans acquired during different sessions. The overall coefficients of repeatability for the test object were between 0.2% and 1.1% and between 1% and 2% for the control group. The range of normal retinal thickness in terms of the 5th and 95th percentiles was 222 to 248 microm in women and 234 to 257 microm in men. CONCLUSIONS: Measurements made from OCT scans were found to be accurate and precise. Introducing water or glycerin into the test object resulted in considerable degradation of the signal, but measurements of gap thickness were still shown to be accurate, precise, reproducible, and repeatable. Retinal thickness measurements in the macular area were repeatable and reproducible. This demonstrates that OCT is a useful tool in the monitoring of patients with conditions that affect macular thickness, even when there is considerable degradation of the OCT signal.
机译:目的:评估由汉弗莱光学相干断层扫描(OCT)系统(汉弗莱-蔡司医疗系统,加利福尼亚州圣莱安德罗)进行的测量的准确性,准确性,可重复性和可重复性。方法:首先通过扫描包含已知尺寸气隙的测试对象来研究系统的性能。间隙中用水或甘油重复测量。在临床环境中,黄斑厚度测量是从20名正常受试者的对照组中获得的。为了进行分析,将这些扫描分为八个部分,每个部分包含10个A扫描。结果:发现所有情况下的平均间隙厚度均接近真实值。会话间再现性的总体系数对于测试对象小于1%,对于对照组,小于1.51%。在不同会话期间获得的扫描之间没有显着差异。测试对象的总体可重复性系数在0.2%至1.1%之间,对照组为1%至2%之间。女性的正常视网膜厚度范围为第5和第95个百分位数,男性为2​​22至248微米,男性为2​​34至257微米。结论:通过OCT扫描进行的测量被发现是准确和精确的。在测试对象中引入水或甘油会导致信号质量明显下降,但间隙厚度的测量仍然显示出准确,精确,可重现和可重复的特点。在黄斑区域的视网膜厚度测量是可重复的和可再现的。这表明即使在OCT信号明显下降的情况下,OCT仍可用于监测患有影响黄斑厚度的疾病的患者。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号