The effects of peripapillary atrophy on the diagnostic ability of stratus and Cirrus OCT in the analysis of optic nerve head parameters and disc size

机译:视神经乳头参数和椎间盘大小分析中,视乳头周围萎缩对层和Cirrus OCT诊断能力的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose. We compared the diagnostic ability of Stratus and Cirrus optical coherence tomography (OCT) in optic nerve head (ONH) analysis, and examined the effects of optic disc size and peripapillary atrophy (PPA) on their diagnostic capacity. Methods. Stratus and Cirrus OCT was performed in 28 control and 78 glaucomatous eyes. ONH parameters and diagnostic capacity calculated from the area under the receiver operating characteristics curve (AUC) were compared between the two modalities. Glaucomatous eyes were classified by optic disc size and the presence/absence of PPA, and their AUCs were compared. Results. Rim area (AUC 0.936) and rim volume (AUC 0.824) showed the best diagnostic capacity in Cirrus and Stratus OCT, respectively. Cirrus OCT showed greater diagnostic ability over Stratus OCT for all ONH parameters. With increasing ONH size, diagnostic ability declined in rim and disc areas, while it improved in average cup-to-disc ratio (CDR), vertical CDR, and cup volume for both modalities. Optic disc size as measured by Stratus OCT was significantly larger than that by Cirrus OCT in glaucomatous eyes with PPA. In addition, the diagnostic capacity of all ONH parameters declined significantly for glaucomatous eyes with PPA, especially for Stratus OCT. In Cirrus OCT, retinal nerve fiber layer (RNFL) thickness showed poorer diagnostic ability than rim area in the glaucoma with PPA group. Conclusions. The diagnostic capacity of Cirrus OCT was superior to that of Stratus OCT. Considering the principle of determining the disc margin, as Stratus OCT tends to make excessive measurements of the optic disc in PPA, caution is required in analyzing the results.
机译:目的。我们比较了Stratus和Cirrus光学相干断层扫描(OCT)在视神经乳头(ONH)分析中的诊断能力,并检查了视盘大小和乳头周围萎缩(PPA)对它们的诊断能力的影响。方法。在28例对照和78例青光眼中进行了Stratus和Cirrus OCT。在两种方式之间比较了从接收器工作特性曲线(AUC)下面积计算出的ONH参数和诊断能力。青光眼按视盘大小和是否存在PPA进行分类,并比较其AUC。结果。边缘区域(AUC 0.936)和边缘体积(AUC 0.824)分别在Cirrus和Stratus OCT中显示出最佳的诊断能力。对于所有ONH参数,Cirrus OCT都比Stratus OCT具有更高的诊断能力。随着ONH尺寸的增加,边缘和椎间盘区域的诊断能力下降,而这两种模式的平均杯碟比(CDR),垂直CDR和杯体积均得到改善。在PPA的青光眼中,由Stratus OCT测量的视盘大小明显大于Cirrus OCT。此外,对于患有PPA的青光眼,特别是对于Stratus OCT,所有ONH参数的诊断能力均显着下降。在Cirrus OCT中,PPA组青光眼的视网膜神经纤维层(RNFL)厚度显示出比边缘区域差的诊断能力。结论Cirrus OCT的诊断能力优于Stratus OCT。考虑到确定光盘余量的原理,由于Stratus OCT倾向于对PPA中的光盘进行过多的测量,因此在分析结果时需要谨慎。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号