Toward the interpretation of the combined effect of size and body weight on the tribological performance of total knee prostheses

机译:解释大小和体重对全膝关节假体的摩擦学性能的综合影响

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: The research questions of the present study were: (1) Is total knee prosthesis wear behaviour influenced by implant size, body weight and their combined effect? (2) Are these findings significant and helpful from a clinical point of view? Methods: Two very different sizes of the same total knee prosthesis (TKP), previously tested with ISO 14243 parameters, were tested on a knee simulator for a further two million cycles using a modified ISO 14243 load waveform. Roughness examination was performed on the metallic components. Gravimetric and micro-Raman spectroscopic analyses were carried out on the polyethylene inserts. Results: The average volumetric mass loss was 69 ± 3 mm3 and 88 ± 4 mm3 for smaller and bigger size, respectively. Bigger TKPs are little influenced by an increased load, while the wear trend of the smaller TKP showed a redoubled slope, and more significant morphology changes were observed. However, the two sizes seem to behave similarly when subjected to a load increase of 15%; the slope of the volumetric mass loss trend was comparable for the two sets of inserts, which did not appear significantly different also at the molecular level. Roughness average parameters of the lateral femoral condyle support this evidence. Conclusions: It can be asserted that the body weight and implant size are relevant to the understanding of TKP wear behaviour. A post-implantation body weight increase in a patient with smaller knee dimensions could results in more critical effects on prosthesis long-term performance.
机译:目的:本研究的研究问题是:(1)膝关节假体的整体磨损行为是否受植入物尺寸,体重及其综合作用的影响? (2)从临床角度来看,这些发现是否有意义且有帮助?方法:使用修改后的ISO 14243负载波形,在膝盖模拟器上测试了之前用ISO 14243参数测试过的两种尺寸完全不同的同一全膝关节假体(TKP),进一步进行了200万次循环。对金属部件进行粗糙度检查。对聚乙烯插入物进行了重量分析和显微拉曼光谱分析。结果:较小和较大尺寸的平均体积质量损失分别为69±3 mm3和88±4 mm3。较大的TKP受负载增加的影响很小,而较小的TKP的磨损趋势则显示出更大的斜率,并且观察到了更为明显的形态变化。但是,当负载增加15%时,这两种尺寸似乎表现出相似的效果。两组插入物的体积质量损失趋势的斜率相当,在分子水平上也没有显着差异。股外侧lateral的粗糙度平均参数支持该证据。结论:可以断定体重和植入物大小与对TKP磨损行为的了解有关。膝关节较小的患者植入后体重增加可能对假体的长期性能产生更关键的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号