Press-fit stability of uncemented hemispheric acetabular components: a comparison of three porous coating systems.

机译:非骨水泥半球形髋臼组件的压配合稳定性:三种多孔涂层系统的比较。

获取原文
获取原文并翻译 | 示例
       

摘要

Cementless acetabular components require good initial fixation to allow bony in-growth. The initial press-fit stability, important for designs that do not rely on supplemental fixation, was examined for three designs with different porous coating systems: beads, fiber mesh, and plasma spray. The ability to withstand tangential loads was determined (maximum rim loads: 122-1730 N). The plasma-sprayed acetabular cups withstood the greatest tangential load within the specified range of motion of 150 microm (P<0.01). Differences in surface preparation of titanium acetabular cups may significantly affect the initial stability of the implants in rim loading.
机译:非骨水泥髋臼组件需要良好的初始固定,以允许骨向内生长。对于不依赖补充固定的设计非常重要的初始压配合稳定性,已针对具有不同多孔涂层系统的三种设计进行了检查:珠,纤维网和等离子喷涂。确定了承受切向载荷的能力(最大轮辋载荷:122-1730 N)。等离子喷涂髋臼杯在指定的150微米运动范围内承受了最大切向载荷(P <0.01)。髋臼钛杯的表面制备差异可能会严重影响植入物在边缘加载时的初始稳定性。