A series of 199 total hip arthroplasties was performed using a porous-coated, hemispherical press-fit acetabular cup. At a mean follow-up of 91.5 months 158 cups were available for clinical and radiological review. The mean age of the patients at the time of the index arthroplasty was 62.5 years. The mean Harris Hip score at final follow-up was 87.3. No shells were revised although eccentric polyethylene wear prompted liner replacement in two cases. Osteolysis was noted in six cases but predominantly in relation to the femoral stem. Focal pelvic osteolytic lesions were rare. All the cups were classified as stable on radiography.
机译:使用多孔涂层,半球形压入式髋臼杯进行了199次全髋关节置换术。平均随访91.5个月,可获得158杯用于临床和放射学检查。进行指数置换时患者的平均年龄为62.5岁。在最终随访中,Harris Hip平均得分为87.3。尽管在两种情况下偏心聚乙烯磨损提示更换衬里,但没有修改外壳。六例发现骨溶解,但主要与股骨干有关。局灶性骨盆溶骨性病变罕见。所有的杯子在放射线照相上被分类为稳定的。