Foot loading with an ankle-foot orthosis: The accuracy of an integrated physical strain trainer

机译:脚踝矫形器的脚负荷:集成式物理应变训练器的准确性

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: To investigate the value of a built-in physical strain trainer for the monitoring of partial weight bearing with an ankle-foot orthosis. Methods: 12 healthy volunteers were asked to perform three trials. Plantar peak pressure values from normal gait (trial one) were defined as 100% (baseline). The following trials were performed with the Vacoped? dynamic vacuum ankle orthosis worn in a neutral position with full weight bearing (trial two) and a restriction to 10% body weight (BW) (trial three), as monitored with an integrated physical strain trainer. Peak plantar pressure values were obtained using the pedar? X system. Results: Peak pressure values were statistically significantly reduced wearing the Vacoped? shoe with full weight bearing for the hindfoot to 68% of the baseline (normal gait) and for the midfoot and forefoot to 83% and 60%, respectively. Limited weight bearing with 10% BW as controlled by physical strain trainer further reduced plantar peak pressure values for the hindfoot to 19%, for the midfoot to 43% of the baseline and the forefoot to 22% of the baseline. Conclusions: The Vacoped? vacuum ankle orthosis significantly reduces plantar peak pressure. The integrated physical strain trainer seems unsuitable to monitor a limitation to 10% BW adequately for the total foot. The concept of controlling partial weight bearing with the hindfoot-addressing device within the orthosis seems debatable but may be useful when the hindfoot in particular must be off-loaded.
机译:目的:研究内置式物理应变训练器对于监测踝足矫形器的部分负重的价值。方法:12名健康志愿者被要求进行三项试验。正常步态(试验一)的足底峰值压力值定义为100%(基线)。以下试验是使用Vacoped?动态负重性踝关节矫形器,穿着中性位,负重(试验2),体重限制在10%(体重)(试验3),并通过集成的物理应变训练器进行监测。足底压力峰值是通过使用踏板获得的。 X系统。结果:戴上Vacoped?后足至基线(正常步态)的68%,中足和前足的足重的分别为83%和60%。由物理应变训练器控制的BW为10%的有限负重可将后脚的足底峰值压力值进一步降低至19%,中脚的基线峰值压力降低至基线的43%,前脚的基线峰值降低至22%。结论:Vacoped?真空踝关节矫形器可显着降低足底峰值压力。集成的物理应变训练器似乎不适合将整个脚的体重限制限制在10%以内。在矫形器内用后足寻址装置控制局部负重的概念似乎值得商but,但在特别是必须卸载后足时可能很有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号