首页> 外文期刊>International Orthopaedics >Long-term results of uncemented acetabular cups with an ACS polyethylene liner. A 14-16-year follow-up study.
【24h】

Long-term results of uncemented acetabular cups with an ACS polyethylene liner. A 14-16-year follow-up study.

机译:使用ACS聚乙烯内衬的髋臼杯未固定的长期效果。一项14-16年的随访研究。

获取原文
获取原文并翻译 | 示例
           

摘要

We assessed the clinical and radiographic results of 40 porous-coated acetabular cups with an Acetabular Cup System polyethylene liner over a minimum 14-year follow-up. Femoral head penetration was estimated using a software package. Fifteen cups were revised, 11 due to polyethylene liner rupture. All cups but two were radiographically stable, and 11 hips showed acetabular osteolysis. The overall femoral head penetration rate in hips without liner fracture with reference to the early penetration point was 0.1188+/-0.070 mm per year. Polyethylene liner fractures were associated with higher early femoral head penetration (P<0.0001) and a vertical cup position (P=0.0016). The 14-year survival without cup revision for any reason was 63.9%, 71.8% with no ACS polyethylene liner fracture and 65.3% with no acetabular osteolysis. Most cups showed a good clinical outcome in general, but major Acetabular Cup System liner failure and osteolysis were frequent. Patients with the ACS cups still in place should be monitored closely.
机译:我们在至少14年的随访中评估了40具髋臼杯聚乙烯衬垫的多孔涂层髋臼杯的临床和影像学结果。使用软件包评估股骨头穿透力。修改了15个杯子,其中11个由于聚乙烯内胆破裂而被修改。除了两个杯子以外,所有杯子均在放射学上稳定,并且11个髋部显示髋臼骨溶解。相对于早期穿刺点,未发生内衬骨折的髋关节整体股骨头穿刺率为每年0.1188 +/- 0.070 mm。聚乙烯内衬骨折与较高的早期股骨头穿透力(P <0.0001)和垂直杯位置(P = 0.0016)相关。不因任何原因而无需改杯的14年生存率为63.9%,无ACS聚乙烯内衬骨折的为71.8%,无髋臼骨溶解的为65.3%。一般而言,大多数杯具良好的临床结果,但髋臼杯系统严重的内衬衰竭和骨溶解很常见。 ACS杯仍在原位的患者应密切监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号