首页> 外文期刊>International Orthopaedics >Measuring the cobb angle with the iphone in kyphoses: A reliability study
【24h】

Measuring the cobb angle with the iphone in kyphoses: A reliability study

机译:用iPhone测量后凸鹿的Cobb角:可靠性研究

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose Smartphones have gained widespread use in the healthcare field to fulfill a variety of tasks. We developed a small iPhone application to take advantage of the built-in position sensor to measure angles in a variety of spinal deformities. We present a reliability study of this tool in measuring kyphotic angles. Methods Radiographs taken from 20 different patients' charts were presented to a panel of six operators at two different times. Radiographs were measured with the protractor and the iPhone application and statistical analysis was applied to measure intraclass correlation coefficients between both measurement methods, and to measure intra- and interobserver reliability Results The intraclass correlation coefficient calculated between methods (i.e. CobbMeter application on the iPhone versus standard method with the protractor) was 0.963 for all measures, indicating excellent correlation was obtained between the CobbMeter application and the standard method. The interobserver correlation coefficient was 0.965. The intraobserver ICC was 0.977, indicating excellent reproductibility of measurements at different times for all operators. The interobserver ICC between fellowship trained senior surgeons and general orthopaedic residents was 0.989. Consistently, the ICC for intraobserver and interobserver correlations was higher with the CobbMeter application than with the regular protractor method. This difference was not statistically significant. Conclusion Measuring kyphotic angles with the iPhone application appears to be a valid procedure and is in no way inferior to the standard way of measuring the Cobb angle in kyphotic deformities.
机译:目的智能手机已在医疗保健领域得到广泛使用,以完成各种任务。我们开发了一个小型iPhone应用程序,以利用内置的位置传感器来测量各种脊柱畸形中的角度。我们介绍了此工具在测量后凸角时的可靠性研究。方法从20个不同患者的图表上拍摄的X射线照片在两个不同的时间呈现给由六名操作员组成的小组。使用量角器和iPhone应用程序对射线照片进行测量,并进行统计分析以测量两种测量方法之间的类内相关系数,并测量观察者间和观察者间的可靠性结果结果计算方法之间的类内相关系数(即,iPhone上的CobbMeter应用程序与标准间的使用量角器的所有测量方法均为0.963,表明CobbMeter应用程序与标准方法之间具有极好的相关性。观察者之间的相关系数为0.965。观察者内部ICC为0.977,表明所有操作者在不同时间的测量结果均具有出色的可重复性。受过研究金培训的高级外科医生与普通骨科住院医师之间的观察者间ICC为0.989。一致地,使用CobbMeter应用程序的观察者内和观察者间相关性的ICC高于常规量角器方法。这种差异在统计学上不显着。结论使用iPhone应用程序测量后凸角似乎是有效的方法,并且绝不逊于后凸畸形中测量Cobb角的标准方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号