首页> 外文期刊>International Orthopaedics >Failure rate of a rotating hinge knee design due to yoke fracture of the hinged tibial insert: a retrospective data analysis and review of the literature.
【24h】

Failure rate of a rotating hinge knee design due to yoke fracture of the hinged tibial insert: a retrospective data analysis and review of the literature.

机译:旋转铰链膝关节设计由于胫骨铰链insert骨骨折而导致的失败率:回顾性数据分析和文献回顾。

获取原文
获取原文并翻译 | 示例
           

摘要

Rotating hinge knee prostheses are known to provide inherent stability. Yoke fractures of the hinged tibial insert of modern generation rotating hinge devices are a matter of continued concern. The aim of this study was to describe incidence and management of yoke fracture of the LPS? hinged tibial insert.Retrospective data analysis of two institutions identified 40 patients with a LPS? total knee arthroplasty. Implant survival and prosthetic complications was calculated according to Kaplan-Meier.Out of the group of 40 patients, four fractures of the metal yoke occurred in four cases (failure rate: 10%). Furthermore, a second fracture occurred in two patients. The overall revision-free prosthetic survival was 57% at 38 months, while prosthetic survival until yoke fracture was 86% at 38 months.Handling yoke fractures as mechanical complication includes replacing the hinged insert, stabilization of the joint and joint line height preservation in order to decrease the cantilever effect at the insert-base plate interface.
机译:已知旋转铰链膝关节假体可提供固有的稳定性。现代旋转铰链装置的铰接胫骨插入物的轭骨折一直是人们关注的问题。这项研究的目的是描述LPS骨骨折的发生和处理。两家机构的回顾性数据分析确定了40例LPS患者?全膝关节置换术。根据Kaplan-Meier计算种植体存活率和假肢并发症。40例患者中,4例发生了4例金属轭铁断裂(失败率:10%)。此外,第二例骨折发生在两名患者中。 38个月时无假体的整体修复存活率为57%,而直到y骨骨折的假体存活率为38%,在38个月时。作为机械并发症处理Hand骨骨折包括更换铰链插入物,稳定关节和保持关节线高度以减少在插入物-基板界面处的悬臂效应。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号