首页> 外文期刊>International Orthopaedics >Biomechanical analysis of a transiliac internal fixator.
【24h】

Biomechanical analysis of a transiliac internal fixator.

机译:trans内固定器的生物力学分析。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: We evaluated the biomechanical characteristics of the transiliac internal fixator (TIFI) as compared to two well-established methods of internal posterior pelvic ring fixation. METHODS: Six freshly frozen human pelves were used for simulated single-leg stance loading of an AO type C injury model (pubic symphysis diastasis and unilateral sacroiliac joint disruption). The symphysis rupture was stabilized with a dynamic compression plate. Afterwards the three internal stabilization systems (TIFI, iliosacral screws and ventral plate osteosynthesis) were analysed. Fragment movement was measured in a contact-free manner with a stereophotometric infrared system. RESULTS: No significant differences in the three-dimensional deformation tolerated by the TIFI as compared to the other internal fixation systems were found. CONCLUSIONS: The transiliac internal fixator provides the same biomechanical stability as the other reference implants tested. We suggest the use of this device as a suitable alternative to the other implants.
机译:目的:我们比较了两种成熟的骨盆后环内固定方法与经皮内固定器(TIFI)的生物力学特性。方法:使用六个新鲜冷冻的人骨模拟AO型C损伤模型(耻骨联合分离和单侧sa关节破坏)的单腿姿态加载。用动力加压板稳定了骨sym的破裂。之后,分析了三个内部稳定系统(TIFI,ili骨螺钉和腹板接骨术)。片段运动是通过立体光度红外系统以非接触方式进行的。结果:与其他内固定系统相比,TIFI耐受的三维变形没有发现显着差异。结论:经internal骨内固定器与其他经过测试的参考植入物具有相同的生物力学稳定性。我们建议将此设备用作其他植入物的合适替代品。