Tibiofemoral contact areas and pressures in six high flexion knees.

机译:六个高屈膝的胫股接触面积和压力。

获取原文
获取原文并翻译 | 示例
       

摘要

The tibiofemoral articulating interfaces of six high flexion knee designs were examined using a standard testing protocol developed by Harris et al. [J Biomech 32:951-958 (1999)] to investigate the polyethylene insert contact areas and pressures. A load of 3600 N was applied for 10 s at 0, 30, 60, 90, 110, 135 and 155 degrees of flexion. Contact areas and pressures at the femoral-polyethylene insert interface were measured with a I-scan 4000 system. Up to 110 degrees of flexion, the VANGUARD RP HI-FLEX showed the highest contact area and lowest pressure. At the deep flexion angles, contact area decreased and contact pressure increased significantly in all knees. The NexGen series showed a constant contact area throughout the various flexion angles. In general, all high flexion knees could result in almost point contact in an extremely high range of motion.
机译:使用Harris等人开发的标准测试方案检查了六个高屈曲膝关节设计的胫股关节接口。 [J Biomech 32:951-958(1999)]研究聚乙烯插入物的接触面积和压力。在0、30、60、90、110、135和155度弯曲度下施加3600 N的载荷10 s。用I-scan 4000系统测量股骨-聚乙烯插入物界面的接触面积和压力。高达110度的弯曲度,VANGUARD RP HI-FLEX表现出最高的接触面积和最低的压力。在深屈曲角度下,所有膝盖的接触面积均减少,接触压力显着增加。 NexGen系列在各种屈曲角度均显示出恒定的接触面积。通常,所有高屈膝都可能导致极高运动范围内的几乎点接触。