A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.

机译:一种新颖的基于Hartman Shack的地形系统:与Scheimpflug + Placido地形图仪和旋转棱镜自动角膜屈光器配合使用,可重复性和一致性。

获取原文
获取原文并翻译 | 示例
       

摘要

The purpose of this study is to analyze the repeatability and agreement of corneal power using a new Hartman type topographer in comparison to Scheimpflug+Placido and autorefractor devices. In this cross sectional, observational study performed at the cornea services of a specialty hospital, 100 normal eyes (100 consecutive candidates) without any previous ocular surgery or morbidity except refractive error were evaluated. All candidates underwent three measurements each on a Full gradient, Hartman type topographer (FG) (iDesign, AMO), Scheimpflug+Placido topographer (SP) (Sirius, CSO) and rotating prism auto-keratorefractor (AR) (KR1, Nidek). The parameters assessed were flat keratometry (K1), steep keratometry (K2), steep axis (K2 axis), mean K, J 0 and J 45. Intra-device repeatability and inter-device agreement were evaluated. On repeatability analysis, the intra-device means were not significantly different (ANOVA, p > 0.05). Intraclass correlations (ICC) were >0.98 except for J 0 and J 45. In terms of intra-measurement standard deviation (Sw), the SP and FG groups fared better than AR group (p < 0.001, ANOVA). On Sw versus Average plots, no significantly predictive fit was seen (p > 0.05, R (2) < 0.1 for all the values). On inter-device agreement analysis, there was no difference in means (ANOVA, p > 0.05). ICC ranged from 0.92 to 0.99 (p < 0.001). Regression fits on Bland-Altman plots suggested no clinically significant effect of average values over difference in means. The repeatability of Hartman type topographer in normal eyes is comparable to SP combination device and better than AR. The agreement between the three devices is good. However, we recommend against interchanging these devices between follow-ups or pooling their data.
机译:本研究的目的是与Scheimpflug + Placido和自动验光仪相比,使用新型Hartman型地形图仪分析角膜屈光力的可重复性和一致性。在该横断面中,在一家专科医院的角膜服务处进行的观察性研究评估了100例正常眼睛(连续100例候选者),除了屈光不正以外,以前没有进行过任何眼科手术或发病。所有候选人均在全梯度,Hartman型地形图仪(FG)(iDesign,AMO),Scheimpflug + Placido地形图仪(SP)(Sirius,CSO)和旋转棱镜自动偏光折射仪(AR)(KR1,Nidek)上进行了三个测量。评估的参数为平面角膜曲率法(K1),陡峭角膜曲率法(K2),陡轴(K2轴),平均值K,J 0和J45。评估了设备内部的可重复性和设备间的一致性。在重复性分析中,装置内平均值无显着差异(ANOVA,p> 0.05)。除J 0和J 45外,类内相关性(ICC)> 0.98。就测量内标准偏差(Sw)而言,SP和FG组的表现优于AR组(p <0.001,ANOVA)。在Sw与平均值图上,未见明显的预测拟合(对于所有值,p> 0.05,R(2)<0.1)。在设备间一致性分析中,均值没有差异(ANOVA,p> 0.05)。 ICC为0.92至0.99(p <0.001)。在Bland-Altman图上的回归拟合表明,平均值对均值差异无临床显着影响。 Hartman型地形图仪在正常眼睛中的可重复性与SP组合设备相当,并且优于AR。这三个设备之间的协议很好。但是,我们建议不要在随访之间交换这些设备或合并其数据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号