Application of a fluoride selective electrode to the stoichiometric analysis of lithium carbonate for nuclear purposes

机译:氟化物选择性电极在用于核目的碳酸锂的化学计量分析中的应用

获取原文
获取原文并翻译 | 示例
           

摘要

An improved version of a potentiometric method for the determination of lithium has been applied to the accurate, high precision analysis of lithium carbonate, a raw material for the solid-state synthesis of lithium metazirconate (Li2ZrO3). This is one of the most promising tritium breeders in fusion reactors. The analytical procedure consisted of a decomposition of the sample in hydrochloric acid followed by evaporation to dryness. The residual lithium chloride was then dissolved in ethanol and titrated with ammonium fluoride in water-ethanol (1 + 1). A lanthanum fluoride solid-state ion-selective electrode was used as end point detector. By increasing the analyte concentration and by dissolving the titrant in the mixed solvent, more accurate results were achieved with respect to the older procedure. A comparison of the results obtained by this method and those obtained by ion-exchange showed no significant difference (P = 0.05). The precision in terms of relative standard deviation was 0.3-0.4%. (c) 2006 Elsevier B.V. All rights reserved.
机译:用于锂测定的电位计方法的改进版本已应用于碳酸锂的精确,高精度分析,碳酸锂是固态合成偏锆酸锂(Li2ZrO3)的原料。这是聚变反应堆中最有前途的tri育种者之一。分析程序包括将样品在盐酸中分解,然后蒸发至干。然后将残留的氯化锂溶解在乙醇中,并用氟化铵的水-乙醇溶液(1 +1)滴定。氟化镧固态离子选择电极用作终点检测器。通过提高分析物浓度并将滴定剂溶解在混合溶剂中,相对于较旧的方法,可以获得更准确的结果。通过这种方法获得的结果与通过离子交换获得的结果的比较显示没有显着差异(P = 0.05)。相对标准偏差的精确度为0.3-0.4%。 (c)2006 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号