During embolization of a large frontal arteriovenous malformation (AVM), Onyx-18 (eV3) was injected into an M3 branch of the middle cerebral artery via a Marathon microcatheter (eV3). After 40 minutes of embolization, the microcatheter could not be retracted due to fixation within the Onyx cast despite prolonged, robust attempts. A balloon microcatheter (Hyper-form TM, eV3) was advanced distally and inflated to provide distal counter tension, allowing microcatheter retrieval with minimal traction on the vasculature.
机译:在大额额动静脉畸形(AVM)栓塞过程中,通过马拉松微导管(eV3)将Onyx-18(eV3)注入大脑中动脉的M3分支。栓塞40分钟后,尽管进行了长时间,可靠的尝试,但由于固定在Onyx石膏中,微导管无法缩回。将球囊微导管(Hyper-form TM,eV3)向远侧推进并充气以提供远侧反张力,从而使微导管收回时对脉管系统的牵引最小。