Validation of a new handheld device for measuring resting metabolic rate and oxygen consumption in children.

机译:用于测量儿童静息代谢率和氧气消耗的新型手持设备的验证。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The purpose of this study was to assess the validity and reliability of the MedGem device to measure resting metabolic rate (RMR) in children. Subjects included 59 children (29 boys, 30 girls; mean age, 11.0 +/- 0.2 y). Subjects were given 4 RMR tests during 1 test session, cconsisting of 2 Douglas bag and 2 MedGem tests, in random counterbalanced order. No significant differences were found between Douglas bag and MedGem systems for oxygen consumption (209 +/- 5 and 213 +/- 5 mL/min, respectively, P = 0.106, r = 0.911, mean +/- standard deviation absolute difference 3.72 +/- 17.40 mL/min) or RMR (1460 +/- 39 and 1477 +/- 35 kcal/d, P = 0.286, r = 0.909, mean +/- standard deviation absolute difference 17.4 +/- 124 kcal/d). Standard error of estimates for oxygen consumption and RMR were 17.4 mL/min and 124 kcal/d, respectively. In conclusion, these data indicate that the MedGem is a reliable and valid system for measuring oxygen consumption and RMR in children.
机译:这项研究的目的是评估MedGem设备测量儿童静息代谢率(RMR)的有效性和可靠性。受试者包括59名儿童(29名男孩,30名女孩;平均年龄11.0 +/- 0.2岁)。在1个测试阶段中,对受试者进行了4次RMR测试,包括2个Douglas袋和2个MedGem测试(以随机平衡的顺序)。 Douglas bag和MedGem系统之间的耗氧量没有显着差异(分别为209 +/- 5和213 +/- 5 mL / min,P = 0.106,r = 0.911,平均值+/-标准差绝对值3.72 + /-17.40 mL / min)或RMR(1460 +/- 39和1477 +/- 35 kcal / d,P = 0.286,r = 0.909,平均值+/-标准偏差绝对差17.4 +/- 124 kcal / d) 。耗氧量和RMR估算值的标准误分别为17.4 mL / min和124 kcal / d。总之,这些数据表明MedGem是一种用于测量儿童耗氧量和RMR的可靠且有效的系统。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号