Clinical feasibility of using an EPID in CINE mode for image-guided verification of stereotactic body radiotherapy.

机译:在CINE模式下使用EPID进行立体定向身体放疗的图像引导验证的临床可行性。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To introduce a novel method for monitoring tumor location during stereotactic body radiotherapy (SBRT) while the treatment beam is on by using a conventional electronic portal imaging device (EPID). METHODS AND MATERIALS: In our clinic, selected patients were treated under a phase I institutional review board-approved SBRT protocol for limited hepatic metastases from solid tumors. Before treatment planning multiple gold fiducial markers were implanted on the periphery of the tumor. During treatment the EPID was used in cine mode to collect the exit radiation and produce a sequence of images for each field. An in-house program was developed for calculating the location of the fiducials and their relative distance to the planned locations. RESULTS: Three case studies illustrate the utility of the technique. Patient A exhibited a systematic shift of 4 mm during one of the treatment beams. Patient B showed an inferior drift of the target of approximately 1 cm from the time of setup to the end of the fraction. Patient C had a poor setup on the first day of treatment that was quantified and accounted for on subsequent treatment days. CONCLUSIONS: Target localization throughout each treatment beam can be quickly assessed with the presented technique. Treatment monitoring with an EPID in cine mode is shown to be a clinically feasible and useful tool.
机译:目的:介绍一种新颖的方法,通过使用常规的电子门成像设备(EPID),在立体定向放射治疗(SBRT)期间监测肿瘤的位置。方法和材料:在我们的诊所中,根据I期机构审查委员会批准的SBRT方案,对部分患者进行了针对实体瘤的有限肝转移的治疗。在制定治疗计划之前,将多个金基准标记物植入肿瘤周围。在治疗期间,EPID以电影模式使用,以收集出射辐射并为每个场产生一系列图像。开发了一个内部程序来计算基准点的位置及其与计划位置的相对距离。结果:三个案例研究说明了该技术的实用性。患者A在其中一个治疗束期间表现出4 mm的系统偏移。从设置时间到分数结束时,患者B的目标漂移约1 cm。患者C在治疗的第一天就出现了不良情况,已量化并在随后的治疗日进行了解释。结论:可以通过提出的技术快速评估整个治疗光束中的靶标定位。以电影模式的EPID进行治疗监测显示是一种临床上可行且有用的工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号