PURPOSE: To verify the usefulness of our developed beam ON-LINE positron emission tomography (PET) system mounted on a rotating gantry port (BOLPs-RGp) for dose-volume delivery-guided proton therapy (DGPT). METHODS AND MATERIALS: In the proton treatment room at our facility, a BOLPs-RGp was constructed so that a planar PET apparatus could be mounted with its field of view covering the iso-center of the beam irradiation system. Activity measurements were performed in 48 patients with tumors of the head and neck, liver, lungs, prostate, and brain. The position and intensity of the activity were measured using the BOLPs-RGp during the 200 s immediately after the proton irradiation. RESULTS: The daily measured activity images acquired by the BOLPs-RGp showed the proton irradiation volume in each patient. Changes in the proton-irradiated volume were indicated by differences between a reference activity image (taken at the first treatment) and the daily activity-images. In the case of head-and-neck treatment, the activity distribution changed in the areas where partial tumor reduction was observed. In the case of liver treatment, it was observed that the washout effect in necrotic tumor cells was slower than in non-necrotic tumor cells. CONCLUSIONS: The BOLPs-RGp was developed for the DGPT. The accuracy of proton treatment was evaluated by measuring changes of daily measured activity. Information about the positron-emitting nuclei generated during proton irradiation can be used as a basis for ensuring the high accuracy of irradiation in proton treatment.
机译:目的:验证安装在旋转龙门端口(BOLPs-RGp)上的已开发的束在线正电子发射断层扫描(PET)系统对剂量-体积-输送-引导的质子治疗(DGPT)的实用性。方法和材料:在我们工厂的质子处理室中,构造了BOLPs-RGp,因此可以安装平面PET装置,其视场覆盖射束辐照系统的等角点。在48例头,颈,肝,肺,前列腺和脑肿瘤患者中进行了活性测量。在质子辐照后的200 s内,使用BOLPs-RGp测量活性的位置和强度。结果:BOLPs-RGp采集的每日活动图像显示每个患者的质子辐照量。质子辐照量的变化由参考活性图像(在第一次治疗中拍摄)与每日活性图像之间的差异表示。在头颈治疗的情况下,在观察到部分肿瘤减少的区域中活性分布改变。在肝治疗的情况下,观察到坏死肿瘤细胞中的清除作用比非坏死肿瘤细胞中的清除作用慢。结论:BOLPs-RGp是为DGPT开发的。质子治疗的准确性是通过测量每日测量活动的变化来评估的。关于在质子辐照期间产生的正电子发射核的信息可以用作确保在质子处理中辐照的高精度的基础。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号