首页> 外文期刊>International journal of pediatric otorhinolaryngology >Cuffed endotracheal tubes in infants and children: A technique to continuously measure the intracuff pressure
【24h】

Cuffed endotracheal tubes in infants and children: A technique to continuously measure the intracuff pressure

机译:婴幼儿气管插管:一种连续测量袖带内压力的技术

获取原文
获取原文并翻译 | 示例

摘要

Objectives: Various devices have been used to measure the intracuff pressure (CP) of an endotracheal tube at the time of inflation; however, no device has found widespread acceptance for the continuous monitoring of CP. We devised a simple method to continuously measure the CP using an invasive pressure monitoring setup (IPMS), which is used routinely in the operating room to monitor arterial or central venous pressures. The accuracy of the device was compared to those obtained from a commercially available and clinically used manometer (MM). Methods: Size 4.0, 5.0 and 6.0mm ID cETTs were placed into one of 3 sizes of polyvinylchloride (PVC) tubes. The cuff of the cETT was inflated after inserting the cETT into the PVC pipes. After inflation, the CP was simultaneously checked using the MM and the IPMS. A total of 100 simultaneous (IPMS and MM) readings were obtained from each of the 3 sizes of cETT with the intracuff pressure randomly varying between 10 and 40cmH2O. Statistical analysis included a Bland-Altman comparison to determine the bias, mean, and 95% levels of agreement (LOA), and a linear regression analysis. Results: Linear regression analysis demonstrated an R2 value of 0.988, 0.9899, and 0.9879 when comparing the pressure from the IPMS and MM from the 4.0, 5.0 and 6.0mm cETT, respectively. Conclusions: A standard IPMS can be used to continuously monitor intracuff pressure with the use of a cETT. Although various techniques can be used safely to avoid hyperinflation of the cuff, they provide only an instantaneous measure of the CP. The current technique allows for continuous monitoring of CP during prolonged cases or in other clinical scenarios when such monitoring is required.
机译:目的:已经使用了各种装置来测量充气时气管导管的袖带内压力(CP)。但是,尚未发现任何设备可以持续监控CP。我们设计了一种简单的方法,使用有创压力监测装置(IPMS)连续测量CP,该装置通常在手术室中用于监测动脉或中央静脉压力。将该设备的精度与从市售的临床压力计(MM)获得的精度进行了比较。方法:将尺寸为4.0、5.0和6.0mm的ID cETT放入3种尺寸的聚氯乙烯(PVC)管之一中。将cETT插入PVC管后,将cETT的袖带充气。充气后,使用MM和IPMS同时检查CP。从cETT的3种尺寸中的每一种中获得总计100个同时(IPMS和MM)读数,袖带内压力在10和40cmH2O之间随机变化。统计分析包括进行Bland-Altman比较以确定偏差,均值和95%一致性水平(LOA),以及线性回归分析。结果:线性回归分析表明,当分别比较IPMS和MM的4.0、5.0和6.0mm cETT的压力时,R2值为0.988、0.9899和0.9879。结论:标准IPMS可用于通过cETT连续监测袖带内压力。尽管可以安全地使用各种技术来避免袖带过度充气,但它们仅提供了CP的即时测量。当前的技术允许在长时间的情况下或在其他需要监测的临床情况下对CP进行连续监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号